SECG 5.0 AIO | User Manual
ECG Special Add-on Parameters
AAMI EC 13 Drift Test:
(ECG 2-27 waveform only): when checked adds 4mVpp 0.1Hz triangle waveform to the ECG
signal (for testing baseline drift).
Respiration:
Provide resistant method and wave modulation method to simulate the change during
human breathing. The system is able to adjust respiration rate, ratio and apnea time.
Resistant Method
To simulate the changes of human breathing via variable resistance. The change is based on
a basic resistance level with variations. Frequency of resistance variation is the same as the
preset respiration frequency.
Wave Modulation Method
Baseline
→
Simulate the human breathing changes via baseline changes of ECG waveform.
Amplitude
→
Simulate the human breathing changes via the amplitude of ECG R wave.
Frequency
→
Simulate the human breathing changes via ECG heart rate frequency.
Apena Time: SECG will output the testing signal for 1 minutes. The testing signal shall stop
after the preset apena time parameter and then repeat the sequence.