MECG 2.0 Test System | User Manual
26
Note : The frequency range is 0.1~100 Hz, the
frequency resolution is 0.1Hz/step.
7.6 Command Replay
To increase test efficiency
, MECG provides “Command Replay” feature. Edit
your test sequences with command script, and automatically output the setting
waveforms.
The command script could be automatically generated while the user operates
MECG. This means user could operate the test procedure for the one time,
then edit, save and replay the script.
Supported Command Scripts:
Table 4: Supported Command Scripts
Command
Parameter
Description
Example
CTS_CSE_Load
CTS/CSE ID
Record code
for the
waveform
Load CTS or CSE
waveform with
optional noise
pattern.
CTS_CSE_Load
PCTH033.CYC
50HZ.N10
Noise
Noise
pattern
LoadLocal_Physionet
Waveform
file path
Waveform
file (*.hea)
path on your
local PC
Load the Physionet
waveform (*.hea)
file on local PC.
LoadLocal_Ph
ysionet
D:\Whaleteq\
WhaleteqMECG
\100.hea
AcceptLeads
LeadList
The lead list
for mapping,
number of
the list
should
match
loaded
waveform.
Map the leads of the
loaded Physionet
waveform.
AccepetLeads
I
II
None
None
None
None V1 V2 V3
V4 V5 V6
LoadLocal_AHA
Waveform
file path
Waveform
file (*.ecg)
path on your
local PC
Load the AHA
waveform (*.ecg)
file on local PC.
LoadLocal_AH
A
D:\whaleteq\
WhaleteqMECG
\1001.ecg
LoadLocal_AHA_TXT
Waveform
file path
Waveform
file (*.txt)
path on your
local PC
Load the AHA text
format waveform
(*.txt) on local PC.
LoadLocal_AH
A_TXT
D:\WhaleteqM
ECG\1001.txt
LoadLocal_CSE
Waveform
file path
Waveform
file (*.DCD)
path on your
local PC
Load the CSE
waveform (*.DCD)
file on local PC.
LoadLocal_CS
E
D:\WhaleteqM
ECG\MA1_001.
DCD