MECG 2.0 Test System | User Manual
25
Waveforms can be started or stopped at
anytime.
․
The output can be started from a mid-
point in the file with “Play (Graph)” button,
but adjusting the starting point “Graph
Start (s)” of the graphs prior to pressing
play.
․
“Next CTS” button, allowing quick access
to the next waveform in the CTS
database
․
A “Looping” checkbox, allowing the user
to repeatedly play the same waveform.
․
“Noise” checkbox, allowing the user to
quickly see the impact of the noise on the
ECG under test
7.5 Other Waveforms
Basic waveforms are selectable as below:
Figure 18: Other Waveforms
The following other functions can be selected:
-
Sine wave, adjustable amplitude and
frequency
-
Triangle wave, adjustable amplitude and
frequency
-
100ms pulse, adjustable amplitude and
frequency
The user needs to press the “load” button to put
these waveforms in memory.
For these settings, the sampling rate is fixed at
1kHz. Due to the relatively low sample rate, this
output is not suitable for frequency response
analysis and should be used for reference only.