![WHALETEQ DFS360 Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/whaleteq/dfs360/dfs360_user-manual_967654026.webp)
DFS360
|
User Manual
25
Item
Name
Description
R
Remove
Remove existing waveforms.
S
Edit
Edit existing waveform
’s name and file
.
T
Save
Save the newly-added or modified waveform
data to the DFS360
/
DFS360D.
The saved waveforms will be automatically
added to the ECG output options. Click Add
button at the lower-left corner in the main
screen to bring up the ECG Output window.
Note:
1.
Users can save up to 10 customized waveforms (max. 100 seconds per
waveform). When the waveform is set for more than 100 seconds, it will
be automatically cut off when loaded.
2.
Up to 20 characters can be set for the waveform name.
3.
If the amplitude of the waveform is set for over 5mV, it will
automatically be scaled to 5mV; if not, then it will remain the same.
4.
The waveform automatically offsets to 0mV.
5.
Support only WhaleTeq format when loaded.
DFS360 AP supports text file (*.txt) only and file content describes as below:
File Format:
[sampling frequency]
[number of samples per signal]
[number of signals]
[signal description (signal-1)] ...
[sample data-1 (signal-1)] ...
[sample data-2 (signal-1)] ...
...
[sample data-N (signal-1)] ...
Description:
sample frequency
Input the frequency of sample data
number of samples per signal