Disassembling and reassembling
1. Hold down the tip ejector.
2. Place the tooth of the opening tool
between the tip ejector and the tip
ejector collar to release the locking
mechanism.
3. Carefully release the tip ejector and
remove the ejector collar.
4. Place the wrench end of the opening
tool over the tip cone and turn
clockwise.
5. After removing the tip cone, wipe the
piston and the tip cone with ethanol
and a lint free cloth.
6. Before replacing the tip cone, grease
the piston slightly using the supplied
silicon grease.
7. After reassembling, use the pipette
several times to ensure that the
grease is spread evenly.
8. Check the calibration of the pipette.