Autoclaving
The pipettes are fully autoclavable by steam at 121°C, 1.05 bar for 15 min.
Autoclave the complete pipette without any disassembling.
After autoclaving under the mentioned conditions, allow the pipette to cool and dry
for 6 hours before use. It is not required to re-calibrate the pipette after autoclaving.
If the pipette is autoclaved frequently, the piston and springs should be greased with
the lubricant supplied with each pipette to maintain smooth movement.
Calibration
Performance testing should take place in a draught-free, stable environment with a
relative humidity above 50% and constant (±0,5 °C) temperature between 15°C and 30°C.
The weighing vessel, pipette, tips, and distilled water need to have been in the room for a
sufficient time, at least 2 hours, to reach equilibrium with the room conditions.
Use an analytical balance with a readability of 0.01 mgs. 10 measurements for each
test volume shall be carried out. For variable volume pipettes, at least three volumes
shall be tested (nominal volume, 50% of nominal volume, and 10% of nominal volume).
1. Select the test volume. Do not change the setting during the test cycle of
10 measurements.
2. Fit the selected tip to the pipette cone.
3. Fill the tip with distilled water and condition the pipette before testing by aspirating
and dispensing the distilled water in the tip five times. Afterwards discard the tip.
4. Attach a new tip on the pipette cone and pre-wet the tip once.
5. Aspirate the distilled water and pipette it into the weighing vessel (use forward
pipetting technique).
6. Weigh the pipetted quantity with an analytical balance and record the weight.
Repeat the test cycle until 10 measurements have been recorded.
7. Convert the recorded masses to volumes, calculate the mean volume, systematic
error and random error as described in EN ISO 8655-6:2002, part 8.
8. Compare the systematic error (inaccuracy) and the random error (imprecision)
with the values in the specification table.
The calibration of the pipette must be set even if only one of the results falls outside
the permitted range.
Recalibration
1. Place the calibration tool into
the holes of the calibration
adjustment lock
(under the push button).
2. Turn the adjustment lock
anticlockwise to decrease and
clockwise to increase
the volume.
3. Repeat checking calibration until
the pipetting results are correct.