Pipetting recommendations
•
Aspirate liquid into the pipette only when a tip is attached. Hold the pipette vertically
when aspirating and dip the tip only 2-3mm into the liquid.
•
Pre-rinse the tip five times before aspiration by filling and emptying the tip.
•
Always control the push button movements with the thumb for consistency.
Allow the liquid, tip and pipette to equilibrate to the ambient temperature before use.
•
When not in use, place the pipette on a bench to avoid transferring body heat
resulting in incorrect dispensing volume
•
Use the correct pipette tip designed for the use with the particular pipette.
•
Select the correct pipetting technique (e.g. reverse, forward pipetting) depending
on the nature of the liquid.
•
Don‘t lay down the pipette with the filled tip, otherwise the liquid can flow into the
pipette and contaminate it
General operation
Volume setting
1. The volume is shown in the handgrip display window.
2. The desired volume is set by turning the push button clockwise or anticlockwise.
ATTENTION!
Using excessive force to turn the push button may jam the mechanism
and damage the pipette.