5. Installation

Location

The unit must be placed on a smooth, level, and sturdy work surface, preferably near a 

drain for easier emptying. Suitable for use in ambient temperatures 5°C to 40°C with a 

maximum humidity of 80% [temperature 31°C] decreasing to 50% [temperature 40°C]. 

The mains voltage fluctuations are not to exceed ± 10°% of the nominal supply.

Unpacking

1. Remove the product from its packaging. 

ATTENTION:

 All the packaging should be kept in safe storage for at least two years for  

warranty purposes.

1.  Remove all protective vinyl coatings prior to use. 
2.  Place the lid onto the bath. Fit the concentric ring sets. 
3.  Fit the power lead into the socket at the rear.

ATTENTION:

 Do NOT touch any electrical contacts or open any closure panels!  

There is a risk of electrical shock.

Power lead and connection to the electrical supply

Ensure that the electrical supply is compatible with the rating label. If in doubt, consult 

a qualified electrician. Where the mains supply or plug connection differs, refer to local 

regulations or consult a qualified electrician.

ATTENTION:

 The product must be earthed!

6. Operation

Connecting constant level and maintaining liquid level

1.  Connect one outlet of the constant level to the water supply using laboratory 

tubing. Adjust the water flow to gently top up the water level whilst maintaining 

boiling. Once connected, the water should flow out of the spout into the bath. At a 

glance, this allows the user to check that the water levels are being maintained.

2.  The second outlet acts as an overflow to a sink or drain. Connect the laboratory 

tubing to suit.

ATTENTION: 

Ensure that the constant level is used correctly to avoid the bath boiling dry. 

Please be aware that the non-resetting safety cut-out is NOT covered under warranty. 