45
5. Measurement Functions and Ranges
DC Current Measuring
Range Resolution
Accuracy
40
µ
A 0.01
µ
A
400
µ
A 0.1
µ
A
4000 µA
1
µ
A
± 1 % v. M. ±3 St.
40 mA
10
µ
A
400 mA
100
µ
A
± 1.5 % v. M. ±3 St.
10 A
10 mA
± 2.5 % v. M. ±5 St.
- Maximum Input Current: 400 mA DC or 400 mA AC in the µA- and
mA-Range (Fuse 0.5 A), 10 A DC or AC in the 10 A-Range (Fuse 10
A)
AC Current Measuring
Range Resolution
Accuracy
40
µ
A 0.01
µ
A
400 µA
0.1
µ
A
4000 µA
1
µ
A
± 1.5 % v. M. ±5 St.
40 mA
10
µ
A
400 mA
100 µA
± 1.8 % v. M. ±5 St.
10 A
10 mA
± 3.0 % v. M. ±7 St.
- Maximum Input Current: 400 mA DC or 400 mA AC in the µA- and
mA-Range (Fuse 0.5 A), 10 A DC or AC in the 10 A-Range (Fuse 10
A)
- The given values apply for frequencies of 50 to 60 Hz