![Westfalia 64 56 89 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/westfalia/64-56-89/64-56-89_instruction-manual_963339019.webp)
14
Operation
Please Notice:
•
When taking measurements in the range under 2 M
Ω
it can take several
seconds before a stable value is reached.
•
If the measuring range is exceeded, a “1” is displayed as overflow
indicator.
•
Keep the measuring tips clean to ensure a good connection.
Accuracy
Range
Resolution
Accuracy
20 K
Ω
1
Ω
± 1 % of Measured Value + 1 Digit
2 M
Ω
100
Ω
± 1 % of Measured Value + 1 Digit
Diode Test Function
With this function you can test diodes and other semiconductors for continuity
and blocking function. The continuity current is 1.2 mA.
Attention!
•
Never connect the measuring tips with a power source while using this
function; this can lead to the destruction of the unit.
•
Only take measurements on circuitry which is disconnected from all power
sources and additionally make sure all capacitors are discharged.
Measuring Procedure
•
Connect the black measuring cable to the COM input (7) and the red
measuring cable to the V/
Ω
input (6).
•
Set the rotary function switch (4) to
.
•
Connect the measuring cables to the component you intend to check and
read the value from the display.
Please Notice:
•
Connect the red measuring cable to the anode (The end of the diode not
marked with a ring) and the black measuring cable to the cathode.