12.
PREVENTIVE MAINTENANCE
12.1. Depending upon the severity of conditions when in use, it is recommended
that you implement a minimum standard preventative maintenance schedule
for the UVC BODYLight and its accessories.
12.2. To maintain compliance for use in Hazardous (Classified) Locations all
components must be in factory original working condition. As such, safety
requires, before and after each use, a thorough visual examination of the unit
and accessories for physical damage that might cause a spark or result in the
ignition of hazardous atmospheres. This examination includes, but is not
limited to, verifying that:
12.2.1. The Light Head of the UVC BODYLight is not cracked, broken, or
breeched.
12.2.2. The Battery Canister for the UVC BODYLight is not cracked, broken, or
breeched.
12.2.3. The Lens is not cracked or chipped.
12.2.4. All Gaskets are in position and not cracked, chipped, or broken.
12.2.4.1.
Ensure that the Gasket is in place around the base of the threads
of the Connector on the Battery Canister, Cord Connected
Battery Pack Manifold, and Cord Connected Power Box.
12.2.4.1.1.
This Gasket is required to be present and functional for
Explosion Proof certification and usage in Hazardous
(Classified) Locations.
12.2.5. The Bumpers are in good condition and capable of protecting the UVC
BODYLight and the work area if struck by the UVC BODYLight.
12.2.6. If using a cord connected Power Box or Battery Pack Manifold, the
Braided Cable Strain Reliefs are not frayed and are gripping the cord
tightly at all points of cord contact, if applicable.
12.2.7. If using a cord connected Power Box or Battery Pack Manifold, the Cord
Jacket is completely intact and that no internal conductors or conductor
insulation is exposed, if applicable.
12.2.8. If using a cord connected Power Box or Battery Pack Manifold, the
Power Box or Battery Pack Manifold is not cracked, broken, or breeched,
if applicable.
12.2.9. If using a cord connected Power Box, the Cord Plug is not cracked or
damaged and the Ground Post is intact, if applicable.
12.2.10. Maintain all Light surfaces, including glass and Light Body, free from
coatings, paint or debris accumulation.