background image

12

Section 4:  TROUBLE-SHOOTING GUIDE

4.10 

Poor Vacuum Readings

 

1. Disconnect vacuum pump from your vacuum system.

 

2. Obtain 1 gallon of DUOSEAL oil, 1 gallon or larger clear plastic container, clear drain hose 

 

 

(1/4 inch ID by at least 2.5 feet long), rubber gloves, goggles, apron and pliers.

 

3. Carry out a forced oil flush (See Section 3.3b) if motor can turn pumping mechanism without 

 

 

excessive  noise.  If pumping mechanism cannot be turned by motor, have pump sent into 

 

 

factory repair center.

 

4. Check ultimate vacuum with thermocouple gauge.  If reading is still unsatisfactory, consult with 

 

 

Welch Technical Service after checking for leaks in your connections.

4.20 

Noisy Pump

 

1.  Locate noise source on your vacuum pump (motor, belt, pump).

 

2.  If motor makes noise:

 

 

. Check for loose bolts on motor mount.

 

 

. Motor may need to be replaced

 

3.  If belt makes noise:

 

 

. Pulley may be bent.  If so, replace.

 

 

. Loose belt, tighten by loosening bolts on motor and moving motor away from pump until belt

 

 

   exhibits a 1/2 inch deflection.

 

4. If pump makes noise: vanes may be sticking, carry out a forced oil flush.

 

5. If noise level is still unsatisfactory, consult with WELCH Technical Service at (847)676-8800, Ext. 1.

4.30 

Excessive Oil Mist

 

1. Check oil level.  Maintain the level between the two lines on the oil window during operation.  

 

     Too high a level will cause excess oil mist.

 

2. Determine the vacuum level in your system with a thermocouple gauge or manometer.  

 

     Be sure the pump is operating at a pressure of 10 Torr or lower for continuous operation.

 

3. A system leak(s) or degassing can cause excessive oil mist.  Locate and seal the vacuum leaks. 

 

     High vapor loads may require a larger pump to compensate.

 

4. An oil eliminator can be installed for coalescing the oil mist from the exhaust gases.  If an oil mist eliminator 

 

     is already installed, the element may need to be replaced if saturated with oil.  

 

     Open up oil mist eliminator and examine the element; replace if needed.

 

5. Check the oil condition for any visible change in color or consistency.  Contaminants in the oil can 

 

     add to the amount of oil exhausted.  Carry out a forced oil flush if contamination is discovered. 

 

     (See Section 3.3b.)

CAUTION

If not, discuss application with Welch Technical service before 

proceedingto prevent oil from overheating and breaking down 

resulting pump failing.

Содержание DUOSEAL 1399B-01

Страница 1: ...sting and causing possible injury to personnel in the area Gardner Denver Thomas Inc Welch Vacuum Technology 5621 W Howard Street Niles IL 60714 Phone 847 676 8800 Fax 847 677 8606 Technical Support 847 588 2366 E Mail gdwelchvacuum gardnerdenver com Part No 67 2063 Web Page www welchvacuum com Printed in USA SAVE THESE INSTRUCTIONS Manufactured in an ISO 9000 Registered Environment 505 872 0037 i...

Страница 2: ...ions are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units confirm to the SI International system of units of measurement The following symbols with recommendation of IEC1010 of warning will be found on the pump Caution refer to accompanying documents Caution risk of electrical shock Caution hot surface WARNING Motor...

Страница 3: ...evel 10 3 30 Shaft Seal Replacement 10 3 40 Repairing Oil Leaks 11 3 50 Repairing Vacuum Leaks 11 3 60 Drive Problems 11 Section 04 Trouble Shooting Guide 4 10 Poor Vacuum Readings 12 4 20 Noisy Pump 12 4 30 Excessive Oil Mist 12 Section 05 Quick Reference Chart Chart 1 13 Chart 2 13 Section 06 Dimenional Drawings with rotation Vacuum Pump 14 Section 07 Pumping Speed Curves Speed Curves 15 Section...

Страница 4: ...y to drive your particular pump at the recommended speed 1 13 Pump Location The pump should be located preferably in a clean and well ventilated area and adequate space should be provided wherever possible for routine maintenance such as changes of oil and belt adjustments and replacements Above all the pump should be located as closely as possible to its system in order to utilize it most efficie...

Страница 5: ...he gas flow is substantially molecular in character The gas flow is then dependent upon the kinetic activity of the molecules to bring it to the intake of the pump 1 162 The Effects of Conductance It has been shown that the conductance of a tube is proportional to the cube of its radius and inversely proportional to its length Therefore it is imperative that the connecting lines be as large in dia...

Страница 6: ...ze yourself with the function and action of the pump which you have now required Remove the intake and exhaust port plugs and temporarily provide a stopper for the intake and a dust cap for the exhaustReview the power requirements as described in Section 1 4 2 103 Cleanliness Take every precaution to prevent foreign from entering the pump A fine mesh screen is provided for this purpose in the inta...

Страница 7: ...of the volume of the mixture this depending upon the amount of air added condensation of the vapor is wither entirely avoided or substantial reduced 2 223 Pump Function With Gas Ballast In a pump functioning on a contaminated system and operating without the gas ballast compression within the stage takes place in the normal manner until the saturation pressure of the contaminating vapor contained ...

Страница 8: ...s rich in hydrogen and observing a sudden change in pressure on an electrical gauge The difference in calibration of these gauges for air and other gases will produce a distinct change in the pressure reading To use this method of detection the system must be under vacuum and the gauge sensing tube must be located between the pump and the area to be probed Use extreme caution as these materials ar...

Страница 9: ... system The required frequency of changes will vary with the particular system Experiences with the process will help you determine the normal period of operation before an oil change is required 3 103 Oil Overheating This pump is designed to operate continuously below 10 Torr Continuous operation of this pump above 10 Torr will lead to overheating and eventual pump failure See Section 2 21 High P...

Страница 10: ... rubber stopper Turn the pump on and run the pump for 10 minutes Close the gas ballast 8 Check the vacuum reading of the pump by connecting a thermocouple gauge tube to the pump s intake If the pump is running nearly as good as when it was new the total pressure reading you will read on wither of these two gauges will be at least 10 micron A simple way to connect the gauge tube to the pump is to r...

Страница 11: ...ing surfaces cleaned of any nicks 3 50 Repairing Vacuum Leaks 3 501 Surface Preparation Good Vacuum seals are an essential and important attribute of a good mechanical pump A good seal is dependent upon the quality of the mating surfaces as well as the sealant and its preparation The mating faces should be carefully inspected for any projections or foreign particles which might interfere with prop...

Страница 12: ...ut a forced oil flush 5 If noise level is still unsatisfactory consult with WELCH Technical Service at 847 676 8800 Ext 1 4 30 Excessive Oil Mist 1 Check oil level Maintain the level between the two lines on the oil window during operation Too high a level will cause excess oil mist 2 Determine the vacuum level in your system with a thermocouple gauge or manometer Be sure the pump is operating at ...

Страница 13: ...02 Wired for 230V 60Hz 3Ph N A Wired for 230V 50Hz 3Ph N A Replacement Parts Shaft Seal Lip Seal 1401E 01 Shaft Seal Mechanical Seal 1401D Air Filter Screen 41 0660 Intake Chamber Cover Gasket 41 0234 Oil Case Gasket 41 0403 Drain Plug 61 2278 Oil Window 41 3753 Large Vane 41 0959 Small Vane N A Large Vane Spring 41 1304 Small Vane Spring N A Large Vane Spring Holder 41 1685 Repair Kits Minor Repa...

Страница 14: ...14 Section 6 DIMENSIONAL DRAWING ...

Страница 15: ...range High RPM direct drive vacuum pumps on the other hand experience a rapid fall off of pumping speed below 100 microns A comparison of the DuoSeal pumping speed curves shown below versus high RPM pump speed curves conclusively illustrates the superior performance of the DuoSeal pump What does that mean for the vacuum pump user DuoSeal pumps provide considerable faster pumpdown time in the low m...

Страница 16: ...16 Section 8 EXPLODED VIEW ...

Страница 17: ...or Repair Kit Catalog 1399K 03 Item 33 exploded view shown on page 18 1401E is a Mechanical Seal Kit which fits all model 1399 serial numbers There is a lipseal on pumps beginning with serial number 25773 and higher It is available as a kit part number 1401E 01 ITEM QTY P N A B DESCRIPTION 20 1 41 1239 1 1 EXHAUST VALVE 21 1 41 1285 1 1 STEEL WASHER FOR SHAFT COLLAR 22 2 41 1304 2 2 VANE SPRING 23...

Страница 18: ...ST FOR PART NUMBER 41 2378 Parts List for Model 41 2378 Item Qty P N Description 1 4 2 01 0308 Screw Hex Hd 5 46 18x1 2 2 2 41 1049 Pump Leg 3 1 41 2088 Oil Case Machined 4 1 41 3753 Oil Case Window 5 1 61 2278 Drain Valve ...

Страница 19: ...on kit Cat No 141601 for models1400 1405 1376 1380 1402 Standard Exhaust Filter Use with pump model Dimension Catalog Number Diameter Height Filter Replacement Element 1392 1399 1400 1376 1380 1402 1400 1397 1374 2 5 5 5 2 5 6 25 9 1417 1417P 10 1417P 20 1417L 1417G 1417H 01 Directional Exhaust Filter Use with pump model Dimension Catalog Number Diameter Height Outlet Tube O D Filter Replacement E...

Страница 20: ...r Height A2 Cat No 1 NW16 to 4 CFM 140 L min 3 75 5 25 8 25 541001 Stainless Steel Element for ISO NW 16 Trap 1 541921 Stainless Steel Element for ISO NW 25 and NW 40 Traps1 541922 Removable 3 quarter center well for dry ice alcohol slurry do not use acetone Visible trapping surface Cold temperature up to 12 hours depending on vapor load Available with straight tube connections ISO Size Diameter H...

Страница 21: ...uum Technology and receipt or possession does not convey any rights to loan sell or otherwise disclose said information reproduction or use of said information for any purpose other than that in connection with said information was supplied may not be made without express written permission of Welch This document is to be returned to Welch upon request and in all events upon completion of the purp...

Отзывы: