2052B-01 PTFE Diaphragm Pump 

MAINTENANCE AND INSTALLATION INSTRUCTIONS 

Read and understand the information and instructions included with this product before using. This information is for your safety and to 
prevent damage to this product. Failure to observe these safety precautions could result in severe bodily injury, including death in extreme 
cases. 

 

CAUTION: To reduce risk of electrical shock… 

1. Do  not  disassemble.  Disassembly  or  attempted  repairs  if  accomplished  incorrectly  can  create  electrical  shock  hazard.    Refer
servicing to qualified service agencies only. 
2. If this pump assembly is supplied with a three-pronged plug, connect unit to a properly grounded outlet only.

 

WARNING: To reduce risk of electrocution... 

1.

This product should never be left unattended when plugged in.

2.

Always unplug this product immediately after use and store in dry place.

3.

Do not use this product in or near area where it can fall or be pulled into water or other liquids.

4.

Do not reach for this product if it has fallen into liquid. Unplug immediately.

5.

To Reduce the Risk of Electric Shock or Injury, Use Indoors Only.

 

DANGER: To reduce risk of explosion or fire... 

1. Do not use this product in or near explosive atmospheres or where aerosol (spray) products are being used.

2

. Do not pump combustible liquids or vapors with this product or use in or near an area where flammable or explosive liquids or

vapors may exist.

3

. Do not use this product near flames.

 

CAUTION: To prevent Injury...

 

1.

Close supervision is necessary when this product is used near children or invalids. Never allow children to operate unit.

2.

Never operate this product if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or
if it has fallen into water. Return the product to a service center for examination and repair.

3.

Keep the cord away from heated surfaces.

4.

Never use while sleeping or drowsy.

5.

Do not operate this product where oxygen is being administered.

6.

This unit is thermally protected. If the motor should overheat, the overload protector will shut the
motor off. If this occurs, turn off the unit and allow the motor to cool until thermal protector resets
automatically.

7.

Always disconnect power source before servicing.

8.

Wear safety glasses or goggles when operating this product.

9.

Use only in well ventilated areas.

10. All  electrical  products  generate  heat.  To  avoid  serious  burns  never  touch  unit  during  or

immediately after operation.

Grounding Instructions 

This product should be grounded. In the event of an electrical short circuit, grounding reduces the risk of electric 
shock  by  providing  an  escape  wire  for  the  electric  current.  This  product  is  equipped  with  a  cord  having  a 
grounding  wire  with  an  appropriate  grounding  plug.  The  plug  must  be  plugged  into  an  outlet  that  is  properly 
installed and grounded in accordance with all local codes and ordinances. 

 DANGER – Improper installation of the grounding plug can result in a risk of electric shock. If repair or 

replacement  of  the  cord  or  plug  is  necessary,  do  not  connect  the  grounding  wire  to  either  flat  blade  terminal. 
The  wire  with  insulation  having  an  outer  surface  that  is  green  with  or  without  yellow  stripes  is  the  grounding 
wire.  Check  with  a  qualified  electrician  or  serviceman  if  the  grounding  instructions  are  not  completely 
understood, or if in doubt as to whether the product is properly grounded. Do not modify the plug provided; if it 
will  not  fit  the  outlet,  have  the  proper  outlet  installed  by  a  qualified  electrician.  This  product  is  for  use  on  a 
nominal  120-volt  circuit,  and  has  a  grounding  plug  that  looks  like  the  plug  illustrated  in  Fig.  A.  A  temporary 
adapter that looks like the adapter illustrated in Fig. B and Fig. C may be used to connect this plug to a 2-pole 
receptacle as shown in Fig. B, if a properly grounded outlet is not available. The temporary adapter should be 
used  only  until  a  properly  grounded  outlet  (Fig.  A)  can  be  installed  by  a  qualified  electrician.

Содержание 2052B-01

Страница 1: ...Operation Manual EN Translation of the german original manual PTFE Diaphragm pump 4 headed Model 2052B 01 2052B 01 UL 2019 08 12 ...

Страница 2: ... by Co Gardner Denver Thomas GmbH We reserve the right to make changes and amendments Gardner Denver Thomas GmbH Am Vogelherd 20 98693 Ilmenau Germany For USA Canada and other Americas Gardner Denver Thomas Inc 1601 Feehanville Drive Mount Prospect IL 60056 USA T 1 847 676 8800 F 1 847 677 8606 welch na gardnerdenver com www gardnerdenver com de de welch T 49 3677 604 0 F 49 3677 604 131 welch eme...

Страница 3: ...ile sleeping or drowsy 5 Do not operate this product where oxygen is being administered 6 This unit is thermally protected If the motor should overheat the overload protector will shut the motor off If this occurs turn off the unit and allow the motor to cool until thermal protector resets automatically 7 Always disconnect power source before servicing 8 Wear safety glasses or goggles when operati...

Страница 4: ...c pressure down to an ultimate vacuum of 2 mbar 1 The pump is equipped with an ON OFF O I power switch 2 Gas Ballast A gas ballast valve A is provided on the pump When condensable vapors are pumped they may be compressed above the saturated vapor pressure and condense in the pump Opening the gas ballast valve admits air into the last stage of the vacuum pump there by reducing the partial pressure ...

Страница 5: ...h to tighten the pressure disc 8 the diaphragm 7 and the strain washer 6 with the correct torque of 5 6 Nm 3 Bring the connecting rod see Fig 3 and the diaphragm 7 into the central position 4 Replace the pump head 5 5 Insert the valves 3 and the O rings 4 Ensure that they are lying completely flat Do not insert the burred side facing the sealing surface Align the connection head flush with the pin...

Страница 6: ...01 is Order No 402042 US 402042 US Service kit consists of Description Kit Order No 402042 US Quantity Order no O Ring ø 12 x 2 9 829217 3 O Ring ø 25 x 2 8 829250 1 Valve 8 400656 Diaphragm 4 400732 Exploded View 2052B 01 Fig 3 Exploded View 2052B 01 ...

Страница 7: ...ot 2 400784 01 13 Rubber pad 1 400785 1 14 Pump head 4 400705 02 15 Connection head 4 410432 16 PTFE insert 4 400902 17 Pressure washer 4 400680 18 Tightening washer 4 400707 19 Diaphragm 4 400732 20 Valve 8 400656 21 Manifold 5 1 400917 01 22 Manifold 2 4 400903 23 Hose nozzle PP DN 8 1x enclosed 2 710798 Hose nozzle PP DN 12 1x enclosed 1 720331 24 Straight threaded joint with seal edge PVDF 10 ...

Страница 8: ...mpleted and returned 2 The product is returned to the factory or other designated service centers freight prepaid 3 The product in our judgment is defective through no action or fault of the user If the product has become defective through misuse abuse or alteration repairs will be billed regardless of the age of the product In this event an estimate of the repair costs will be submitted and autho...