49
Character
Number of
Data
Field Name
Field Description
Number
Characters Type
213
2
uChar
Number of
The number of reflex tests performed.
Reflex Tests
Range=0 to 4
215
2
uChar Reflex Test
Reflex test selection parameters.
Parameters
Bit 0=Ipsi status:
0=Not selected
1=Selected
Bit 1=Contra status: 0=Not selected
1=Selected
Bit 2=500 Hz status: 0=Not selected
1=Selected
Bit 3=1000 Hz status: 0=Not Selected
1=Selected
Bit 4=2000 Hz status: 0=Not Selected
1=Selected
Bit 5=4000 Hz status: 0=Not selected
1=Selected
Bits 6 & 7=Not used
217
2
uChar Reflex #1
Results of Reflex test #1.
Result
Bits 1 & 0=dB level tested:
00=Low dB level
01=Middle dB level
10=High dB level
11=Invalid
Bits 3 & 2=Reflex tests status:
00=NT=Not tested
01=YES=Yes, a reflex was
detected
10=NR=No reflex detected
11=NT_CAL=Not tested,
due to a calibration data error
Bits 5 & 4=Test frequency:
00=500 Hz
01=1000 Hz
10=2000 Hz
11=4000 Hz
Bit 6=Test type:
0=IPSI
1=CONTRA
Bit 7=Not used
219
4
ulnt
Reflex #1Base- Average baseline compliance,
line Average
in cm3 X 256
223
4
ulnt
Reflex #1 Re-
Average reflex compliance,
flex Average
in cm3 X 256
227
4
ulnt
Reflex #1
Reflex compliance data point
Reflex [0]
#1 of 4, in cm3 X 256.
Bit 15=Noise indicator
0=Quiet data measurement
1=Noisy data measurement,
potentially unreliable
:
:
:
:
:
239
4
ulnt
Reflex #1
Reflex compliance data point
Reflex [3]
#4 of 4, in cm3 X 256.
Bit 15=Noise indicator
0=Quiet data measurement
1=Noisy data measurement,
potentially unreliable