Update Propaq CS DFU 3.4X
15
Stop Monitoring a Patient
If you want to discontinue monitoring the patient, follow these steps.
1.
Press the
NET OFF
key on the upper left corner of the monitor display.
(From the Main Menu you can also press
SETUP,
ACUITY
,
END TELE
.)
2.
Disconnect the leads and sensors from the patient and turn off the
power to the monitor.
If you do not turn off power to the monitor within about 30 seconds, the
Wireless Propaq CS will automatically try to reconnect to the network.
If you do not use NET OFF or END TELE to disconnect from the network
as described above, the Acuity Central Station generates a DROPOUT
equipment alert at Acuity.
If you want to monitor this same patient at a later time, you will need to re-
select the patient name from Wireless Propaq CS or confirm the patient ID
at Acuity.