Guidance and manufacturer’s declaration – electromagnetic immunity
IEC 61000-4-11
0% 1 cycle and 70%
25/30 cycles
Single phase: at 0
0% 300 cycle
0% 1 cycle and 70%
25/30 cycles
Single phase: at 0
0% 300 cycle
Power frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration –electromagnetic immunity
The ProBP™ 2000 Digital Blood Pressure Device is intended for use in the electromagnetic environment specified below.
The customer or the user of the ProBP™ 2000 Digital Blood Pressure Device should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz:
3 Vrms
6Vrms (in ISM and
amateur radio
bands)
80% Am at 1kHz
150 kHz to 80 MHz:
3 Vrms
6Vrms (in ISM and
amateur radio
bands)
80% Am at 1kHz
Recommended separation distance
Portable and mobile RF communications equipment
should be used no closer to any part of the ProBP™ 2000
Digital Blood Pressure Device, including cables, than the
recommended separation distance calculated from the
equation appropriate for the frequency of the
transmitter.
Recommended separation distances:
d=0.35;
d=1.2
Radiated RF
IEC 61000-4-3
10V/m,
80% Am at 1kHz
10V/m,
80% Am at 1kHz
80 MHz to 800
MHz:
d=1.2
800 MHz to 2.7
GHz:
d=2.3
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer, and
d
is the
recommended separation distance in
meters (m). Field strengths from fixed
RF transmitters, as determined by an
electromagnetic site survey
a
, should be
less than the compliance level in each
frequency range
b
. Interference may
occur in the vicinity of equipment
marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
Directions for use
EMC guidance and manufacturer’s declarations 33
Содержание ProBP 2000
Страница 1: ...Welch Allyn ProBP 2000 Digital Blood Pressure Device Directions for use Software Version 1 X...
Страница 4: ...iv Contents Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 20: ...16 Maintenance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 26: ...22 Specifications Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 30: ...26 General radio compliance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 32: ...28 Warranty Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 34: ...30 Approved accessories Welch Allyn ProBP 2000 Digital Blood Pressure Device...