Troubleshooting
This section includes a list of error messages and frequently asked questions for
problems you may encounter with your blood pressure device. If the device is not
operating as you think it should, check here before contacting Welch Allyn Technical
Support:
http://www.welchallyn.com/support
.
Problems and error messages
The device presents technical alarms and low-priority physiological alarms. Technical
alarms occur when there is a device-related error. Physiological alarms occur when blood
pressure measurements fall outside of set alarm limits.
Technical alarms
Problem
Symptom
Root cause
Solution
No power
Display will not light up
Batteries are drained.
Replace with new batteries
Batteries are inserted
incorrectly.
Insert the batteries correctly
AC adapter is inserted
incorrectly.
Insert the AC adapter tightly
Low batteries The display indicates the
“BAT-LO” message,
pauses for 3 seconds. The
battery icon shows empty
(does not flash).
Batteries are low.
Replace with new batteries
Error messages E 01 shows
The cuff is not secure.
Readjust the cuff, have the patient relax for a
moment, and then measure again
E 02 shows
The cuff is very tight
Refasten the cuff and then measure again
E 03 shows
There is too much
pressure in the cuff.
Refasten the cuff and then measure again
E 10 or E 11 shows
The device detected
motion while measuring.
Readjust the cuff, have the patient relax for a
moment, and then measure again
E 20 shows
The measurement process
does not detect a pulse
signal
Loosen the clothing on the patient's arm and
then measure again
E 21 shows
The measurement is
incorrect
Have the patient relax for a moment and then
measure again
Visit www.welchallyn.com for further assistance.
14 Maintenance
Welch Allyn ProBP™ 2000 Digital Blood Pressure Device
Содержание ProBP 2000
Страница 1: ...Welch Allyn ProBP 2000 Digital Blood Pressure Device Directions for use Software Version 1 X...
Страница 4: ...iv Contents Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 20: ...16 Maintenance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 26: ...22 Specifications Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 30: ...26 General radio compliance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 32: ...28 Warranty Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Страница 34: ...30 Approved accessories Welch Allyn ProBP 2000 Digital Blood Pressure Device...