WALL MOUNT SET-UP
CAUTION: DO NOT mount
into sheetrock. Must be
mounted into stud.
The LS200 Procedure
Light arm is designed not to go
beyond 90˚ and the working
distance is 24" (60cm). To
ensure proper placement on
the wall, the top edge of the
wall plate should be located
4" (10cm) above the working
surface.
NOTE: Wall plate must be
leveled using a carpenter’s
level to ensure precise, drift-free movement.
Mounting to wood stud:
1. Remove wall mount assembly and hardware kit from box.
2. Select mounting location on a stud in the wall that is within 8'
(2.4m) of electrical outlet.
3. Using wall plate as a template, place on selected location on wall,
level plate and mark both holes.
4. Remove wall plate and drill pilot holes using 0.12" (3mm) drill bit.
5. Place wall plate on wall
and drive top 1.5" (3.8cm)
wood screw
halfway into
wood stud
.
6. Align wall plate using
a level and secure at
bottom with 1.5" (3.8cm)
wood screw.
7. Tighten top wood screw.
8. Prepare for transformer
mounting by following
“Transformer Mounting”
instructions through
step 4 on page 8.
WOOD STUD
LEVEL
WOOD SCREW
5
EXTENSION
ARM (G)
4"
4"