PRETEST TYMP CHECKS
For your convenience, a test cavity is provided with your GSI 38 Auto Tymp. This test
cavity enables you to quickly verify, on a daily basis, the prober calibration of your unit.
GSI strongly recommends that you make this quick check a part of your daily routine.
TYMP CALIBRATION VERIFICATIONS
To initiate the quick check, select the Tymp only mode and insert the probe into the
0.5
cm
3
opening on the test cavity. See Figure 2-l 1.
NOTE
Since the GSI
38
is designed to start automatically, it is
important that the probe is inserted as quickly and as
smoothly as possible. During the calibration check, the
probe must be held carefully and without movement.
Do not place the probe on the same counter as the
instrument or any moving object during this check as
mechanical noise will be picked up by the probe and
interfere with the calibration check.
Содержание GSI 38
Страница 1: ......
Страница 3: ...1 t t L c c k 4 c i I e PRODUCT SPECIFICATION 2...
Страница 14: ...i i L i i c FUNCTIONAL DESCRIPTION 13 c...
Страница 29: ...CIRCUIT THEORY L 28...
Страница 36: ...i L c i k t CALIBRATION t 35...
Страница 51: ...TROUBLE SHOOTING L b L I 4 50...
Страница 57: ...L L L k DISASSEMBLY 56...
Страница 59: ......
Страница 60: ......
Страница 63: ...I L L c t t AUDIO TYMP BOARD ASSEMBLY DRAWINGS SCHEMATICS 62...
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 80: ...PROBE 79...
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 87: ...L L L L PUMP 86...
Страница 88: ......
Страница 89: ...i PRINTER e 88...
Страница 90: ......
Страница 92: ...I I c DISPLAY L L i i 91...
Страница 93: ......
Страница 95: ...I I c ASSEMBLY L L i i 94...