Table 4-l: CALIBRATION REFERENCE THRESHOLDS
Freq
(Hz)
125 250 500 750 1000
1500 2000 3000 4000 6000 8000
dB
45.0 25.5 11.5 8.0 7.0 6.5 9.0 10.0 9.5 15.5 13.0
Standard Reference Threshold Levels re: 20 uPa for Telephonics
TDH39-P earphones as measured on the National Bureau of Standards
9-A coupler. Reference ANSI S3.6-1989, IS0 389-1975 Standards.
4.2
CALIBRATION MODE OPERATION
The Calibration Mode is used to perform the transducer
calibration. This calibration information is used during the
Normal Mode to determine the correct attenuator position to
produce the required SPL for the selected HL. The Calibration
Mode consists of two modes.
The Custom SPL Selection Mode allows the selection of customized
SPL values for alternate transducers. The Transducer SPL
Calibration Mode is used to calibrate the sound pressure level
(SPL) of each stimulus for each transducer.
4.2
.1 CAL/NORMAL SWITCH
This switch is used to select the operating mode of the
instrument. The Calibration Mode is entered by setting this
switch to the CAL position while Dip Switch S1 is in the OFF
position. The Calibration Mode is exited by setting this switch
to the NORMAL position.
4.2.2
CALIBRATION MODE DIP SWITCHES
(See
"Calibration Mode Dip Switch In-Depth Description" later in
this section.)
These Dip Switches select various options of the Calibration
Mode. They are read to determine the selected options when the
Calibration Mode is entered.
Содержание Grason-Stadler GSI 17 1717-9700
Страница 1: ...Covers Welch Allyn AM232 Audiometer ...
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 17: ...R 3 P io R 5 Figure 2 1 CONNECTORS CONTROLS INDICATORS 8 ...
Страница 23: ......
Страница 24: ......
Страница 43: ......
Страница 44: ......
Страница 49: ......
Страница 50: ......
Страница 53: ......
Страница 54: ......
Страница 66: ......
Страница 67: ......
Страница 69: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 95: ...Figure 10 2 BATTERY POWER CONNECTION 88 ...
Страница 98: ... v7044 0037 17 17 0406 Figure 10 3 NiCad BATTERY OPERATION 9 1 ...
Страница 100: ......
Страница 102: ......
Страница 103: ......