CAUTION
Do not disconnect or power off the device during
the license installation.
7. Once you see the
Add new features
screen confirming the license installation, click
Close
.
8. For the license upgrade to take effect, press and hold the power button for about 8
seconds to reboot the device.
(Optional) Upgrade your existing CP 150 device to DICOM
through the Welch Allyn Service Tool administration menu
1. Download the Service Tool Partner Connect and install on your PC.
2. Follow the instructions to set up a user name and password.
3. Plug your CP 150 device into a USB port on your PC running the Service Tool and
turn on the device.
4. Activate the license through the Administration menu. If the Welch Allyn Service
Tool is not still open, go to the Windows Start menu, and then select
All Programs
> Welch Allyn > Welch Allyn Service Tool
.
5. Log in to the Welch Allyn Service Tool.
6. Click on
Administration > Install license
.
7. Use the radio button to select
Medical device license
.
8. Enter the code you received for additional DICOM features for your device into the
authorization code field.
9. To choose the CP150 device, highlight the CP150 in the Device list.
10. Click
Select
.
Service manual
Service 43
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......