Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Keyboard Assembly (CP 100)
Procedure
1.
Open unit using the
directions.
2.
Disconnect the flat cable from the
keyboard.
3.
Remove the 13 screws holding in the
keyboard assembly.
4.
Gently lift out the keyboard assembly.
Reassembly Note
1.
Install and tighten the 3 screws first
(shown in
) to align the keyboard.
DO NOT tighten the screws past the
specified torque.
2.
Install remaining 10 screws using the
specified torque.
Figure 58. Replacing the Keyboard Assembly (CP 100)
Flat Cable
Screws
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...