Service Manual
CP 100 and CP 200 Electrocardiograph
51
6.
Slide the paper tray cover until it locks into
place before lifting the paper tray assembly.
7.
Slowly lift the paper tray assembly and
remove the connections. See
Figure 35. Lifting the Paper Tray Assembly
Reassembly Note
1.
Keep the black printhead harness wires
behind the tabs on the auxiliary tray cover as
you place the cover into the assembly.
2.
When pressing the auxiliary paper cover into
place, listen for two distinct clicks when the
tabs lock into place.
Figure 36.
Procedure
Paper Tray
Assembly
Connections
Printhead
Harness
Ground
Printhead Harness Wires
Tabs
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...