- 33 -
_________________________________________________________________________
2.5 Calibration Procedure: Single Level AudioScope
•
Calibration:
Materials Required: Audiometer Handle Test specification A00277
1 sheet (This is the process.)
Electronics Module Test Specification, A00273
2 sheets (These are the specifications.)
Note
:A note appearing on the specification reads"Exact Measurements are not
required for 100% inspection". This note does not apply to sound settings.
Adjust the AudioScope according to the printed sound settings and their
tolerances as shown on the print.
Your Sequence:
Original Step number from A00277: Modifications
if
any:
•
2.5.6
____ step 1
Start Button
Check
for
smooth
operation.
•
2.5.6
____ step
2
On/Off Switch
•
2.5.7
____ step 3
L.E.D.s
LED.s must be bright.
•
2.5.8
____ step 4
Rattle
•
2.5.9
____ step 7
Cover
•
2.5.10____
step 8
Pneumatic Seal
Check seal before sound.
•
2.5.11____
step 5
Sound
Level
•
2.5.12____
step 9
Charge
•
2.5.13____
step 10
Lamp Ejector
•
2.5.14____
step 11
Trained
Listener
Calibration is complete.
If Single level AudioScope does not take calibration, unit is not serviceable.
Customer should be advised that unit is obsolete, parts are not available .
Lamp, battery, and charging system can be used on new AudioScope 3.
Содержание AudioScope
Страница 2: ......
Страница 6: ...Section 1 General Information...
Страница 24: ...24 Section 2 Service ______________________________________________________________________...
Страница 38: ...Section 3 Troubleshooting...
Страница 41: ...41...
Страница 42: ...42 Section 4 Disassembly and Repair...
Страница 47: ...47 Section 5 Drawings Test Specifications...
Страница 48: ...48...
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......