![Welch Allyn 105313 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/welch-allyn/105313/105313_manual_954048049.webp)
EMC guidance and manufacturer’s
declarations
EMC compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment. This device complies with IEC EN 60601-1-2:2007.
•
All medical electrical equipment must be installed and put into service in accordance
with the EMC information provided in this document and the
Welch Allyn CP 150 12-
lead resting electrocardiograph Directions for Use.
•
Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.
The electrocardiograph complies with all applicable and required standards for
electromagnetic interference.
•
It does not normally affect nearby equipment and devices.
•
It is not normally affected by nearby equipment and devices.
•
It is not safe to operate the electrocardiograph in the presence of high-frequency
surgical equipment.
•
However, it is good practice to avoid using the electrocardiograph in extremely close
proximity to other equipment.
Electromagnetic emissions
The CP 150 electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or user of the CP 150 electrocardiograph should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The CP 150 electrocardiograph uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The CP 150 electrocardiograph is suitable for use in all establishments
other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions Class A
45
Содержание 105313
Страница 1: ...CP 150 12 lead resting electrocardiograph Directions for use...
Страница 21: ...Example ECG report Directions for use ECG home screen 17...
Страница 22: ...18 ECG home screen CP 150 12 lead resting electrocardiograph...
Страница 30: ...26 ECG tests CP 150 12 lead resting electrocardiograph...
Страница 32: ...28 Saved tests CP 150 12 lead resting electrocardiograph...
Страница 36: ...32 Settings CP 150 12 lead resting electrocardiograph...
Страница 48: ...44 Troubleshooting CP 150 12 lead resting electrocardiograph...
Страница 57: ......
Страница 58: ...Reorder No Material No 105313 720134...