Item
Button
1
Leads button
2
Gain button (size)
3
Speed button
4
Filters button
8. If an
Artifact
message appears, minimize the artifact, as described under
Troubleshooting. You might need to ensure that the patient is comfortably warm,
reprepare the patient’s skin, use fresh electrodes, or minimize patient motion.
9. (Optional) Adjust the waveforms, using the buttons to cycle through the following
options:
•
leads displayed
•
gain (size)
•
speed
•
filters
10. Touch the
button to perform the Auto ECG test.
11. If a
Waiting for 10 seconds of quality data
message appears, at least 10 seconds
of ECG data have been collected with excessive artifact. Time requirements in the
message may vary, based upon selected print format. Minimize the artifact, as
described under Troubleshooting. Then wait for the test to print. If necessary, you
can override the wait time and print the available data immediately, but the result
might be an incomplete or poor-quality test.
12. After the test prints, select the desired option:
Reprint
,
Save
, or
Retest
. If the
Auto
Save
setting is turned off, touch
Save
to save the test. Select one of the following
locations:
•
Local (internal memory)
•
USB mass storage device (Any tests that you save to a USB mass storage
device can be retrieved only from a CardioPerfect workstation.)
•
Workstation
24 ECG tests
CP 150 12-lead resting electrocardiograph
Содержание 105313
Страница 1: ...CP 150 12 lead resting electrocardiograph Directions for use...
Страница 21: ...Example ECG report Directions for use ECG home screen 17...
Страница 22: ...18 ECG home screen CP 150 12 lead resting electrocardiograph...
Страница 30: ...26 ECG tests CP 150 12 lead resting electrocardiograph...
Страница 32: ...28 Saved tests CP 150 12 lead resting electrocardiograph...
Страница 36: ...32 Settings CP 150 12 lead resting electrocardiograph...
Страница 48: ...44 Troubleshooting CP 150 12 lead resting electrocardiograph...
Страница 57: ......
Страница 58: ...Reorder No Material No 105313 720134...