36
Maintenance
EN
8. Maintenance
8.1 Intervals
We recommend only having
servicing and maintenance
performed by Weinmann, the
manufacturer, or authorized
dealers.
Both filters must be regularly checked for soiling.
• The coarse filter should be cleaned once a week
and changed every 6 months.
• The fine filter should be changed after 250 oper-
ating hours at the latest.
From reasons of hygiene we recommend:
• replacing the pressure measurement tube
6 months and earlier when soiled;
• replacing the complete mask system every 6 to 12
months depending on soiling;
• replacing the exhalation system according to the
relevant operating instructions.
In addition, servicing must be carried out as a preventative
measure.
• After every 5000 operating hours (maintenance
symbol
appears in display);
• 2 years at the latest (see servicing sticker on bot-
tom of unit).
Maintenance includes the following:
• Filter change;
• Cleaning unit;
• Replacement of any defective parts;
• Replacement of battery;
• Complete check of unit functions and pressure
indicator.