Operation
19
EN
Activate automatic system
(from software version 6.0)
A 0
= Automatic system not
activated
A 1
= Automatic system
activated
1. With the SOMNO
smart
2 switched off, press
until the current setting for
softPAP exhalation relief (
soft 1
,
soft 2
or
soft 0
) appears in the display. Finally release
the Softstart button
2. Press humidifier button
briefly while the softPAP
setting is being shown in the display. The current
setting for Auto switch on/off “
A 0
” flashes in the
display.
3. Press the on / off switch
to change the setting
to “
A 1
“.
If a button is not operated for 5 seconds, the display
will disappear and ”
0
” will reappear.
When the automatic system is activated, you can
switch SOMNO
smart
2 on and off using the on / off
switch
or using the pressure available in the mask.
• SOMNO
smart
2 is switched on automatically as
soon as the patient breathes into the mask
(pressure > 0,5 hPa).
• If the mask is removed, the unit will switch off after
5 seconds.
• If an oronasal mask with integrated emergency ex-
halation system is used, the device does not switch
on/off automatically even if setting “A” is selec-
ted. The emergency exhalation valve of the mask
means that the device cannot detect the pressure
change required for switching.