Wavetek 132 Скачать руководство пользователя страница 6

SECTION

OPERATION

INSPECTION

The following procedures should be performed to as-

sure the user that the instrument has arrived at its

destination in satisfactory operating condition. Com-

plete calibration and checkout instructions are pro-

vided in Section 4 to determine compliance with elec-

trical specifications.

Checking Visually

7.

After carefully unpacking the instrument, visually in-

spect the external parts for damage to knobs, dials,

indicators, surface areas, etc. If damage is discovered,

file a claim with the carrier who transported the in-

strument. Retain the shipping container and packing

material for use in case reshipment is required.

Checking Electrically

N O T E

lnstrumen ts are normally shipped con-

nected for 115 V power unless 230 V

power is ordered. Refer to the end of

this section for conversion instructions.

The steps in this paragraph provide a quick checkout

of the instrument operation. If electrical deficiencies

exist, refer to the WARRANTY in the front of this

manual. The following test equipment is recommended

for performing this electrical inspection:

Name

Required

Characteristics

Oscilloscope

Plug-l n

Plug-l n

Counter-Timer

To 30 MHz

Dual channel

Peak mV  measuring

capability

To 2 MHz with 5-digit

resolution

1.

Turn FREQ HZ selector to the Xl K position.

(This connects ac power to the unit and estab-

lishes the frequency multiplier.)

2.

3.

4.

5.

6.

8.

9.

10.

11.

12.

13.

14.

15.

16.

17.

18.

19.

Depress MODE - FUNC push button.

Connect oscilloscope to the

 OUT connec-

tor with a 

 terminator.

Set frequency dial to the 1.0 mark and FREQ

VERNIER to CAL position.

Set function selector to the

 position.

Set OUTPUT ATTEN (dB) and OUTPUT VER-

NIER (dB) to maximum clockwise (cw)  posi-

tion (no attenuation).

Check for 1 kHz sine wave with at least 10 V

p-p amplitude on oscilloscope.

S e l e c t

 and

 with function selector

and check for 10 V p-p amplitude on oscil-

loscope.

Turn frequency dial from maximum counter-

clockwise (ccw) to maximum cw position and

check for frequency change.

Step OUTPUT ATTEN (dB) selector through

its range and verify attenuation at each step.

Rotate OUTPUT VERNIER (dB) control from

maximum cw to maximum ccw position and

check for decreasing amplitude.

Rotate FREQ VERNIER control and check for

frequency change.

Set FREQ VERNIER control at maximum cw

and frequency dial at 0.02. Set frequency to

20 Hz with counter. Connect a 0 to +5 Vdc

input to the VCG IN connector. Slowly increase

voltage input from 0 to +5 V and check that

frequency of output waveform increases from

approximately 20 Hz to 2 kHz.

Depress SEQUENCE LENGTH

- 1 push

button.

Set NOISE FREQ HZ selector to the 16K/160K

position and vernier control fully cw.

Connect a BNC cable from the NOISE SYNC
connector (rear panel) to the external trigger

input of the oscilloscope.

Rotate the function selector to

 and check

to assure oscilloscope displays digital noise.

Check to assure SEQUENCE LENGTH push

buttons vary the bits in the sequence length.

Step NOISE FREQ HZ selector through its

range and check to assure clock frequency

changes. (Use NOISE CLOCK connector on rear

panel.)

3

Содержание 132

Страница 1: ...INSTRUMENT OPERATION AND MAINTENANCE THE INFORMATION IN THIS DOCUMENT MAY NOT BE DUPLICATED IN ANY MANNER WITHOUT THE PRIOR APPROVAL IN WRITING OF WAVETEK SAN DIEGO 9045 Balboa Ave San Diego CA 92123...

Страница 2: ...s can be used program ensures that the latest electronic developments individually or combined to provide selectable cali are incorporated into the Wavetek instruments by the brated signal to noise an...

Страница 3: ...strument comply with installation instructions MAINT ENANCE may requir e power on w i th th e instrumen t cover sremoved shoul d be don eonly by qualified personne l awar eof the elect h a zards WARNI...

Страница 4: ...circuitry is capable of 1000 1 deviation The VCG amplifier has a 100 kHz bandwidth and a slew rate of 0 1 The instantaneous frequency is the result of the sum of the dial setting and the externally a...

Страница 5: ...he amount of fre quency deviation NOTE When noise is added to the signal output specifications apply up to 200 kHz and the square wave rise time is dera ted by a factor of 10 In the clock range of 1 6...

Страница 6: ...UNC push button Connect oscilloscope to the OUT connec tor with a terminator Set frequency dial to the 1 0 mark and FREQ VERNIER to CAL position Set function selector to the position Set OUTPUT ATTEN...

Страница 7: ...nd noise generators This control has no affect on the noise frequency 2 FREQ VERNIER Allows precision electronic control of the signal output frequency A full turn of the control is approximately equa...

Страница 8: ...pen circuit providing 20 V peak to peak maxi mum or into a load providing a 10 V peak to peak output OUTPUT ATTEN dB Attenuates the output both signal and noise from 0 dB to 60 dB in six calibrated 10...

Страница 9: ...he value to the right of the detent mark VCG IN This connector allows external volt age control of function generator frequency Up to 1000 1 frequency change may be obtained A positive voltage increas...

Страница 10: ...termination Select the desired waveform by setting function 1 2 3 4 5 Terminate OUT connector with 50 ohm 2 watt termination Select the desired waveform by setting function selector to or Set FREQ HZ...

Страница 11: ...rce 1 Select the desired signal waveform by setting function selector to or 2 Set frequency dial and FREQ HZ range multi plier for desired signal output frequency 3 Depress MODE S N push button 4 Sele...

Страница 12: ...t output signal amplitude by setting OUT PUT ATTEN dB control to appropriate atten uation position and fine adjusting signal to de sired amplitude with OUTPUT VERNIER dB control Operation as a Digital...

Страница 13: ...lt use Converting a 115 volt unit for 230 volt operation is a simple matter 2 Loosen two captive thumb screws on rear panel and remove panel 1 Remove power cord 3 4 The conversion switch is located on...

Страница 14: ...hes function like a Schmitt trigger with the limit points set at the wave form extremes firing when the triangle wave reaches 1 25 volts and 1 25 volts When firing occurs the hysteresis and output swi...

Страница 15: ...is considerably less in amplitude than the triangle wave input and is thus amplified to be equal to the triangle wave Either square triangle or sine waveforms can be se lected as a signal source The n...

Отзывы: