Wavetek 132 Скачать руководство пользователя страница 11

6.

b. For increasing frequency sweep with positive

dc input, set dial to lower frequency limit.

c. For decreasing frequency sweep with nega-

tive dc input, set dial to upper frequency

limit.

To sweep the audio range from 20 Hz to 20  kHz,

set the controls to 20 Hz as follows:

a. Set the main dial to 0.02.

b. Set the frequency vernier to the full ccw

position.

c. Introduce a 0 to +5 V ramp into the VCG

inputconnector.

The maximum frequency which may be obtained, when

using the VCG input, is the range multiplier times the

value at the top of the dial (2.0). The minimum fre-

quency is the range multiplier times the value at the

bottom of the dial (0.002 - frequency vernier must

be in ccw position). Verify this using the nomograph

in Figure 2-3.

MAIN DIAL

VCG IN

 OUT

SETTING

VOLTAGE

FREQUENCY

.02

.2

.4

.6

.8

1.6

1.8

2.0

A

B

C

Figure 2-3

-

VCG Voltage-to-Frequency Nomograph

In example 1, the dial is set at 1.0 and 0 voltage is

applied to the VCG input. Extend a straight line from

1.0 (dial setting) through 0 voltage (VCG voltage) and

obtain a dial frequency of 1.0. For the total output

frequency, multiply the range by 1 with the same dial

setting. Example 2 shows the results of using a ramp

from -1 volt to +2.5  volts for the VCG voltage. This

results in a swept output from 0.6 to 2.0 on the dial.

Remember to multiply the dial times the range.

Operation as a Calibrated Signal-to-Noise Source

1.

Select the desired signal waveform by setting

f u n c t i o n   s e l e c t o r   t o

,

, or

.

2.

Set frequency dial and FREQ HZ range multi-

plier for desired signal output frequency.

3.

Depress MODE - S/N push button.

4.

Select noise bandwidth by setting NOISE FREQ

HZ control to desired range and fine adjusting

bandwidth by turning the noise frequency Hz

vernier control.

5.

Select desired SEQUENCE LENGTH by de-

pressing appropriate push button.

6.

Select signal-to-noise ratio by setting S/N - N/S

(dB) control to appropriate attenuation position

and fine adjusting attenuation to desired value

with S/N - N/S vernier control.

7.

Select total output amplitude by setting OUT-

PUT ATTEN (dB) control to appropriate atten-

uation position and fine adjusting signal to de-

sired amplitude with OUTPUT VERNIER (dB)

control.

Operation as a Calibrated Noise-to-Signal Source

1.

2.

3.

4.

5.

6.

7.

Select the desired signal waveform by setting

f u n c t i o n   s e l e c t o r   t o

,

, or

,

Set frequency dial and FREQ HZ range multi-

plier for desired signal output frequency.

Depress MODE - N/S push button.

Select noise bandwidth by setting NOISE FREQ

HZ control to desired range and fine adjusting

bandwidth by turning the noise frequency Hz

vernier control.

Select desired SEQUENCE LENGTH by depres-

sing appropriate push button.

Select noise-to-signal ratio by setting S/N - N/S

(dB) control to appropriate attenuation position

and fine adjusting attenuation to desired value
with S/N - N/S vernier control.

Select total output amplitude by setting OUT-

PUT ATTEN (dB) control to appropriate atten-

uation position and fine adjusting signal to de-

Содержание 132

Страница 1: ...INSTRUMENT OPERATION AND MAINTENANCE THE INFORMATION IN THIS DOCUMENT MAY NOT BE DUPLICATED IN ANY MANNER WITHOUT THE PRIOR APPROVAL IN WRITING OF WAVETEK SAN DIEGO 9045 Balboa Ave San Diego CA 92123...

Страница 2: ...s can be used program ensures that the latest electronic developments individually or combined to provide selectable cali are incorporated into the Wavetek instruments by the brated signal to noise an...

Страница 3: ...strument comply with installation instructions MAINT ENANCE may requir e power on w i th th e instrumen t cover sremoved shoul d be don eonly by qualified personne l awar eof the elect h a zards WARNI...

Страница 4: ...circuitry is capable of 1000 1 deviation The VCG amplifier has a 100 kHz bandwidth and a slew rate of 0 1 The instantaneous frequency is the result of the sum of the dial setting and the externally a...

Страница 5: ...he amount of fre quency deviation NOTE When noise is added to the signal output specifications apply up to 200 kHz and the square wave rise time is dera ted by a factor of 10 In the clock range of 1 6...

Страница 6: ...UNC push button Connect oscilloscope to the OUT connec tor with a terminator Set frequency dial to the 1 0 mark and FREQ VERNIER to CAL position Set function selector to the position Set OUTPUT ATTEN...

Страница 7: ...nd noise generators This control has no affect on the noise frequency 2 FREQ VERNIER Allows precision electronic control of the signal output frequency A full turn of the control is approximately equa...

Страница 8: ...pen circuit providing 20 V peak to peak maxi mum or into a load providing a 10 V peak to peak output OUTPUT ATTEN dB Attenuates the output both signal and noise from 0 dB to 60 dB in six calibrated 10...

Страница 9: ...he value to the right of the detent mark VCG IN This connector allows external volt age control of function generator frequency Up to 1000 1 frequency change may be obtained A positive voltage increas...

Страница 10: ...termination Select the desired waveform by setting function 1 2 3 4 5 Terminate OUT connector with 50 ohm 2 watt termination Select the desired waveform by setting function selector to or Set FREQ HZ...

Страница 11: ...rce 1 Select the desired signal waveform by setting function selector to or 2 Set frequency dial and FREQ HZ range multi plier for desired signal output frequency 3 Depress MODE S N push button 4 Sele...

Страница 12: ...t output signal amplitude by setting OUT PUT ATTEN dB control to appropriate atten uation position and fine adjusting signal to de sired amplitude with OUTPUT VERNIER dB control Operation as a Digital...

Страница 13: ...lt use Converting a 115 volt unit for 230 volt operation is a simple matter 2 Loosen two captive thumb screws on rear panel and remove panel 1 Remove power cord 3 4 The conversion switch is located on...

Страница 14: ...hes function like a Schmitt trigger with the limit points set at the wave form extremes firing when the triangle wave reaches 1 25 volts and 1 25 volts When firing occurs the hysteresis and output swi...

Страница 15: ...is considerably less in amplitude than the triangle wave input and is thus amplified to be equal to the triangle wave Either square triangle or sine waveforms can be se lected as a signal source The n...