Wavelength Electronics LD2TC5 LAB Скачать руководство пользователя страница 23

23

TROUBLESHOOTING

PROBLEM

POTENTIAL CAUSES

SOLUTIONS

The LDD output switched off unexpectedly

OR

I can't enable LDD output current

The output disabled due to an interlock. 
Check error messages on screen or with 

ERR?

 query

Fix appropriate interlock or follow 
instructions on Error Messages Page.

TC failure detected.

If the TC output also disabled, resolve 
the source of that error.  If LDD output 
was remotely linked to the TC status, 
but you don't want it to be linked, use 

LASer:ENABle:OUTOFF 

command to 

unlink them.

Temperature limits have been reached.

If the temperature limits can be increased 
without damage, increase the temperature 
limit settings.

The TC output switched off unexpectedly

OR

I can't enable TC output current.

Check error messages on screen or with 

ERR?

 query, OR TC failure was detected.

The instrument detected an error. Follow 
instructions on Error Messages Page.

Temperature limits have been reached.

If the temperature limits can be increased 
without damage, increase the temperature 
limit settings.

The instrument is on but the front panel 
display is blank.

Instrument firmware is being downloaded 

from USB port.

Wait  patiently  until  the  firmware  has 

finished updating.

The screen brightness is too low.

While  the  screen  may  appear  off,  the 

brightness is at a low setting.  Go to the 
settings screen to raise the brightness 
or send remote command (

BRIGHT) 

to 

increase brightness.  If you can't see the 
screen because it is too dark and don't 
want to easily use the remote command, 
imagine the location of the icons on the 
screen. Good luck.

LD noise current is higher than expected.

Ground Loop, bad connections and solder 
joints, noisy computer connections, etc.

Refer to 

AN-LD09 Troubleshooting Low 

Noise Systems.

Actual LD current does not match the 
setpoint.

The setpoint exceeds the limit current 
setting.

If the limit current can be increased without 
damage to the laser, increase the limit 
current setting.

The laser is compliance voltage limited.

If the limit voltage can be increased without 
damage to the laser, increase the limit 
voltage setting. See Additional Technical 
Information for more details.

Temperature does not stabilize very well at 
setpoint.

Poor thermal contact between components 
of the thermal load.

Use thermal paste or washers between the 
load/TEC and TEC/heatsink interfaces. 
Make sure the temperature sensor is in 
good thermal contact with the load.

Operating outside of the ideal region of the 
temperature sensor.

The sensor type and bias current should be 
selected to maximize sensitivity at the target 
temperature. Thermistors provide the best 
performance, particularly for applications 
where a single setpoint temperature must 
be accurately maintained. For example, at 

25°C a 10 kΩ thermistor has a sensitivity 

of 43 mV/°C, whereas an RTD sensor has 
a sensitivity of 4 mV/°C.

Proportional control term is set too high.

Run an IntelliTune scan.

Derivative term is introducing noise.

Run an IntelliTune scan.

Содержание LD2TC5 LAB

Страница 1: ...ale DB15 2 Connector Kit WCB407 TE RH Sensor DB15 2 Cable WCB408 LD Type A B DB15 Cable WCB409 LD Type C DB15 Cable WCB319 TC to LD Mount Cable WCB326 Type A B Laser to LD Mount Cable WCB327 Type C La...

Страница 2: ...rformance and value The touch screen interface makes operation intuitive and simple A LabVIEW virtual instrument remote command set and USB interface allow integration into an instrument library Safet...

Страница 3: ...the following conditions indoor use 6500 ft 2000 m 80 maximum relative humidity for temperatures up to 31 C and decreasing linearly to 50 relative humidity at 50 C transient overvoltages to Overvolta...

Страница 4: ...el this is already done on the connectors delivered with the instrument 2 For the active interlock insert a BNC terminator in the ACTIVE LOCK on the rear panel provided with the instrument 3 On the fr...

Страница 5: ...reen 2 Press the blue text next to PD Biasing to toggle whether the bias voltage for the photodiode is enabled or disabled 3 Press the blue text next to PD Range until the value matches the appropriat...

Страница 6: ...iode Cathode 10 Output Current Monitor 11 Laser Diode Setpoint Monitor 12 Monitor Ground 13 No Connection 14 No Connection 15 No Connection EXT MOD ACTIVE LOCK BNC PINS Active Signal Ground Figure 7 E...

Страница 7: ...nput Range 5 to 5 V Sums with onboard setpoint External Modulation Input Damage Threshold 6 V Current Limit Range 0 to 101 Leakage Current 1 A Measured with unit on and setpoint at zero Current Monito...

Страница 8: ...61000 4 6 IEC 61000 4 8 IEC 61010 1 EN61010 2 30 EN61326 1 European Council Directive 2004 108 EC European Council Directive 2006 95 EC Size 3 86 8 82 17 7 98 224 450 in mm Weight 9 26 4 2 lbs kg POW...

Страница 9: ...ral Derivative PID control loop to adjust the drive current The terms are user adjustable and when properly configured will quickly settle the load temperature with minimal overshoot and ringing Intel...

Страница 10: ...and electrical earth ground 3 Laser Diode Out Sensor Type A B Laser diode connector and sensor input 15 pin female D SUB connector 4 Laser Diode Out Sensor Type C Laser diode connector and sensor inp...

Страница 11: ...when values are set 4 TC Enable Button Allows instrument to deliver current to the thermoelectric load Glows blue when current is enabled 5 LDD Enable Button Allows instrument to deliver current to t...

Страница 12: ...equire heatsinking RLOAD TEC A1 TEC A2 Test Load Thermistor Test Load Figure 11 Test Temperature Load Sensor Wiring Schematic Setting up this test load allows current to be driven to either the positi...

Страница 13: ...nstrument If either of the Enable buttons is lit press the button to disable output current Once all output current is disabled the rear Power switch may be pressed to the O position powering down the...

Страница 14: ...gle through the Profile options select one and then press the Save icon The following operational parameters are saved TC PARAMETERS LD PARAMETERS Tune Mode Control Mode Temperature Setpoint Laser Set...

Страница 15: ...efined tolerance the target icon will be visible NOTE PD Current is displayed here if the PD Transfer Function specified on the Photodiode Screen is zero OFF PD Power is displayed here if the PD Trans...

Страница 16: ...the rear panel and turn on the rear power switch 4 Open the Firmware Update Application 5 Select the unit to update in the application 6 Drag drop the file into the application window or browse for t...

Страница 17: ...will be ringed with red SELECT A SENSOR The sensor choice determines the sensor voltage to temperature relationship and which sensor pins are monitored on the 15 2 D SUB 1 Press to access the Sensor S...

Страница 18: ...tom sensor using remote commands THERMISTOR Displays the Steinhart Hart coefficients Figure 28 Thermistor Calibration Screen RTD Displays the Standard Curve and the resistance at 0 C Figure 29 RTD Cal...

Страница 19: ...ction is chosen for Tune mode Setpoint Response This setting has a fast time to temperature and will overshoot Disturbance Rejection This setting has a quick return to setpoint after stability is reac...

Страница 20: ...is within the window defined by setpoint tolerance FAN CURRENT DRIVE PINS 10 11 This output will drive 12 V up to 0 5 A to power a fan for airflow across the thermoelectric and load LD OPERATING INST...

Страница 21: ...mode the setpoint chosen corresponds to actual current through the photodiode proportional to output power This mode requires the user to configure additional photodiode settings 1 Press the photodiod...

Страница 22: ...1 A V Table 5 Model dependent external modulation transfer functions The input range is always 5 V to 5 V Damage will occur at 6 V input ADDITIONAL CAPABILITIES ANALOG CURRENT MONITOR PINS 10 12 This...

Страница 23: ...RIGHT to increase brightness If you can t see the screen because it is too dark and don t want to easily use the remote command imagine the location of the icons on the screen Good luck LD noise curre...

Страница 24: ...rent is enabled Thermoelectric wiring is reversed Reverse the leads to the thermoelectric The thermoelectric is generating too much heat and has inadequate cooling Install a larger heatsink or a fan t...

Страница 25: ...t to reach setpoint Increase the TC current limits and begin the IntelliTune scan again Temperature going in the wrong direction IntelliTune aborted TEC polarity is reversed Reverse the connection to...

Страница 26: ...TEC or larger heatsink or increase the limit if safe Laser disabled because temperature is below min limit Minimum limit is too high Unlink the Laser output from TC see LASer ENABle OUTOFF or if safe...

Страница 27: ...utput current at the limit and continue operating at that level of output current Output current will not be disabled if the limit is reached INDEPENDENT LINKED OPERATION The LDTC LAB can operate with...

Страница 28: ...NS 16 35 415 3 17 75 450 9 0 60 15 2 0 80 20 3 3 47 88 1 6 0 152 4 3 91 99 3 8 86 225 0 16 35 415 3 3 47 88 1 3 91 99 3 8 86 225 0 Dimensions in inches mm 5 CABLING SPECIFICATIONS WCB407 TE RH SENSOR...

Страница 29: ...29 WCB408 LD TYPE A B DB15 CABLE NOT INCLUDED WCB409 LD TYPE C DB15 CABLE NOT INCLUDED...

Страница 30: ...30 WCB319 TC TO LDMOUNT NOT INCLUDED WCB326 TYPE A B LASER TO LDMOUNT NOT INCLUDED WCB 327 TYPE C LASER TO LDMOUNT NOT INCLUDED...

Страница 31: ...y Reseller are expressly prohibited from reverse engineering decompiling or disassembling this product NOTICE The information contained in this document is subject to change without notice Wavelength...

Отзывы: