Appendix 3.
Static field measurements: WPH-DC
Warning:
• Under high field conditions (above 30 mT), it is not recommended to connect
the
SMP2
directly to the
WPH-DC
probe as the
SMP2
contains ferromagnetic
elements. In these cases, you must use the 5-meter extension cable to keep
the
SMP2
as far away as possible.
Important note:
if it is suspected that the measuring point may perhaps have an
elevated field, it is recommended to carry out a preliminary assessment around that
point, approaching progressively. If it is found that the field is close to 30 mT, where
the
SMP2
will be placed to take the final measurement, precautions should be taken to
prevent the
SMP2
from moving, or using the extension cable as mentioned previously.
• The
WPH-DC
probe integrates hall sensor technology that allows very high
fields to be measured (up to 10 T) but these sensors have a residual voltage
(without the presence of a magnetic field) that must be compensated to
perform an accurate measurement. For this, you would have to use the “zero
Gauss” camera delivered with the equipment and make the zeroing of the
probe. (More detail below).
User’s Manual
Electromagnetic Field Meter SMP2
P.
68