5.2.6 Offline mode
In this mode (checkbox "Send data to control center" unticked), the device will
not make any communication to the Control Centre or send any alarms. Visual
and audible alarms still will work.
The program can be used to download the data stored in the device in TXT
format (see Appendix 2). Click the “Download measurement” button. Data must
be downloaded when the probe used to make the measurements is connected
and it cannot be downloaded when the device is in USB power mode.
5.2.7
Malfunctions
If you have any problem with the program or the device:
• Download the device log by clicking “Download unit Log” (if you use the
MonitEM-Lab
in offline mode).
•
Download the program session file by clicking the
Tools
Save Log
menu.
•
Send these files to your technical support contact.
Note:
if the red light status (SD card) stays on all the time when the USB is connected,
please disconnect the USB and restart the unit.
5.3 MonitEM-Lab Ethernet configuration
MonitEM-Lab
is configured by default to work with DHCP (automatic IP
configuration). In order to know what IP the device has, you can use the
MonitEM Reader
software by going to
File > Connect through Ethernet
.
Note:
First you need to switch on the
MonitEM-Lab
and connect it to the Ethernet.
In this new window you can do a search ("SEARCH" button) of the equipment in
your local network. The search can be automatic or you can define an IP range.
It should show one or more IP corresponding to your equipments. They can be
connected by clicking on "Click to connect" from the list.
User’s Manual
Electromagnetic Radiation Monitoring Device MonitEM-Lab
P.
13