CLX-XT (28652)
Page 1
REV 5 (5/2020)
Specifications
Measurement Range
0.00 – 10.00 mg/L (PPM) or Optional 0.00 – 15.00 Mg/L
Accuracy
Standard 0-10 mg/L range:
±5% of reading or ±0.03 mg/L (PPM) whichever is greater for range of
0-6.0 mg/L(PPM) & ±10% of reading from 6.01-10.00 mg/L (PPM)
Optional 0-15 mg/L range:
±10% of reading or ±0.03 mg/L (PPM) whichever is greater for the
entire range
Resolution
0.01 mg/L (PPM)
Cycle Time
Adjustable; 60 seconds to 10
minutes (600 seconds)
Note: the system defaults to 60 seconds ( most software versions)
Display
Multi-Line Liquid Crystal Backlit Display
Alarms
Two Programmable, 120-240VAC 2A Form C Relay
Analog Output
Powered 4-20 mA, 600
Ω
drive, isolated
Communications Port
Bi-directional RS-485 with Modbus
Water Pressure
Integral pressure regulator 0.34 bar (5.0 PSI) to 10.3 bar (150 PSI.)
Flow Rate to Waste
200 – 400 ml/min.
Operating Temperature
0°C – 55°C (32°F – 131°F)
Wetted Materials
PVC, Borosilicate Glass, Reslyn (FFKM), Viton
®
(FKM),
Polypropylene, 316 Stainless Steel, Acetal, Noryl
®
, Silicone, PPS
Sample Temperature Range
0°C – 55°C (32°F – 131°F)
Power Supply
100
– 240 VAC, 47 – 63 Hz, 250VA
100 -240 VAC ±10% for serial numbers 201705625 or greater
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Environmental Conditions
Not recommended for outdoor use.
Altitude up to 2000 meters
Up to 95 % RH (non-condensing)
Enclosure Rating
Enclosure certified to IP 66,
Fan rated IP 55
Regulatory Compliance
And Certifications
Tested to IMO MEPC. 174(58) Part 3
Reference Imanna Report # 18953-1
CE Approved, Listed to UL 61010-1 Ed 3: 2012
Certified to CAN/CSA C22.2 No. 61010-1: 2012
EMC to EN61326-1: 2013
Shipping Weight
11.8 kg (26lbs.) Reagents are Shipped Separately
Shipping Dimensions
63cm X 64cm X 32cm (24.5” X 25 X 12.5”)