3.  Remove fittings from solenoid valve inlet and outlet.
4.  Attach fitting to new solenoid valve inlet and outlet.
5.  Install reagent tubing to both sides of solenoid valve.
6.  Plug in wiring to harness.
7.  Reinstall two screws removed in step 1.

10.0 Troubleshooting

Symptoms

Possible Cause

Solution

4-20mA output 

repeats 2-3 times. 

(Frozen reading)

Fault Error.

Visually inspect the instrument and 

correct the fault.

No sample water 

intake.

Inlet water line clogged.

Add a filter to the inlet water line.

For technical assistance, please contact the HF scientific Technical Services Department. 

HF scientific

16260 Airport Park Drive, Suite 140

Fort Myers, FL 33913

Phone: 239-337-2116

Toll Free: 888-203-7248

Fax: 239-454-0694

Email: [email protected]

Website: www.hfscientific.com

11.0 Warranty

Watts Regulator Co. (the “Company”) warrants each municipal market instrument product to be free from defects in material and workmanship 

under normal usage for a period of one (1) year from first use or two (2) years from date of the Company’s invoice from the original sale of the 

product, whichever occurs first. In the event of such defects within the warranty period, the Company will, at its option, replace or recondition 

the product without charge. Parts which by their nature are normally required to be replaced periodically, consistent with normal maintenance, 

specifically reagents, desiccant, sensors, electrodes and fuses, are excluded. Also excluded are accessories and supply-type items. 
Proof of purchase from the Company (Company invoice or paid order confirmation) and/or first use (commissioning) must be provided when making 

a product warranty claim.

THE WARRANTY SET FORTH HEREIN IS GIVEN EXPRESSLY AND IS THE ONLY WARRANTY GIVEN BY THE COMPANY WITH 

RESPECT TO THE PRODUCT. THE COMPANY MAKES NO OTHER WARRANTIES, EXPRESSED OR IMPLIED. THE COMPANY HEREBY 

SPECIFICALLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE IMPLIED 

WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

The remedy described in the first paragraph of this warranty shall constitute the sole and exclusive remedy for breach of warranty, and the Company 

shall not be responsible for any incidental, special or consequential damages, including without limitation, lost profits or the cost of repairing or 

replacing other property which is damaged if this product does not work properly, other costs resulting from labor charges, delays, vandalism, 

negligence, fouling caused by foreign material, damage from adverse water conditions, chemical, or any other circumstances over which the 

Company has no control. In addition, the Company shall not be responsible for any costs incidental to the Company’s warranty response efforts, 

including, without limitation, costs associated with the removal and replacement of systems, structures or other parts of facilities, de-installation, 

decontamination and re-installation of products, or transportation of products to and from the Company. This warranty shall be invalidated by 

any abuse, misuse, misapplication, improper installation or improper maintenance of the product, alteration of the product or use of any parts or 

accessories (including but not limited to reagents) not provided by the Company.
Some states do not allow limitations on how long an implied warranty lasts, and some states do not allow the exclusion or limitation of incidental or 

consequential damages. Therefore the above limitations may not apply to you. This Limited Warranty gives you specific legal rights, and you may 

have other rights that vary from State to State. You should consult applicable state laws to determine your rights. 

SO FAR AS IS CONSISTENT 

WITH APPLICABLE STATE LAW, ANY IMPLIED WARRANTIES THAT MAY NOT BE DISCLAIMED, INCLUDING THE IMPLIED 

WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED IN DURATION TO ONE YEAR 

FROM THE DATE OF ORIGINAL SHIPMENT.

IS-HF-CSM-100184 2224

 100184

 

© 2022 HF scientific 

USA:  

T: (239) 337-2116 • Toll-Free (888) 203-7248 • F: (239) 454-0694 • HFscientific.com

Latin America:  

T: (52) 81-1001-8600 • HFscientific.com

Содержание 28062

Страница 1: ...ilver Monitor IS HF CSM 100184 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference Rev 1 ...

Страница 2: ...ing the Units of Measurement 7 5 3 Setting the Date Time 7 5 4 Adjusting the Brightness 7 5 5 Setting the Time Schedule 7 5 6 Setting the Reading Mode 8 5 7 Adjusting the Line Flush Duration 8 5 8 Security Setting 8 5 9 Firmware Update 9 5 10 Copying FRAM data to SD Card 9 5 11 Factory Reset 10 5 12 About Device 10 6 0 System Start Up 10 6 1 Prepare Reagents 10 6 2 Priming the CSM 10 6 3 Taking Re...

Страница 3: ...nt This pictorial alerts you to the need to read the manual This pictorial alerts you to electricity electrocution and shock hazards 1 0 Understanding Safety Information This manual contains safety and use instructions that must be followed during the installation commissioning operation care and maintenance and service of the CSM All responsible personnel must read this manual prior to working wi...

Страница 4: ...o 23 hours 59 minutes Path Length 10mm Display 3 5 Color graphic touch screen Analog Output Powered 4 20mA 1000 Ω drive isolated Maximum Water Pressure 124 kPa 18 psi see Plumbing Connections section Flow Rate 185 ml min to 225 ml min 044 to 059 Gal min Operating Temperature 1 C 50 C 34 F 122 F Max Sample Water Intake and Drainage Flow 225 ml minute Wetted Materials PBT Polypropylene 316 Stainless...

Страница 5: ...be removed for easier connections Please ensure this terminal set is properly tightened using the two side screws 3 2 Plumbing Connections All plumbing connections are made with user supplied OD tubing using the push to connect fittings attached to the CSM All water tubing used to connect to the CSM is supplied by the installer Sample water should be particulate free The sample water supply does n...

Страница 6: ...electrical panel and tighten Securing the Electrical Enclosure Once all electrical connections are made the electrical panel must be secured by closing the compartment and tightening the captive Phillips screw 4 0 Commissioning and Start Up 1 Once all water and electrical connections have been made turn on power to the CSM 2 The CSM will run through a startup process that takes less than 10 second...

Страница 7: ...ve 4 Once the date and time have been saved press the button to return to the home screen or press Cancel to return to the previous menu 5 4 Adjusting the Brightness The screen brightness can be adjusted to provide easier readability of the display in low light or no light conditions There are 10 brightness levels available To adjust the screen brightness 1 Press Brightness on the Configuration Me...

Страница 8: ... suggested line flush duration is 1 second for every foot of sample line tubing assuming inlet pressure of 4 60psi To adjust the line flush duration 1 Press Line Flush Setting on the Configuration Menu screen 2 Using the up and down arrows set the desired duration for the line flush in seconds 3 Once the desired Line Flush Duration has been saved press the button to return to the home screen or pr...

Страница 9: ...ate is complete a restart of the instrument is required Do not remove the SD card or turn power off while the software update is in progress Doing so can cause an unrecoverable failure NOTICE 5 10 Copying FRAM data to SD Card The following information from the CSM can be transferred to an SD card in CSV file format Readings Error and warning logs with date and time stamps All user settings Calibra...

Страница 10: ...he prime sub menu 3 Press the desired priming option If this is the first time starting up the instrument start with Prime Water and repeat this step for all priming options in this menu Reagent Bags Hanging on Brackets P N 110097 Assembled in the USA Reagent Kit Copper Indicator 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 L 74125 Rev E 11 2020 WARNING Harmful if swallowed Causes skin i...

Страница 11: ...per Silver 500 ml Solution C in Cal Procedure DI Water 500 ml Solution D in the Cal Procedure To recalibrate the CSM follow the prompts on the screen 1 From the screen press Service to enter the Service Menu 2 Press CALIBRATE to begin the Calibration sequence Ensure you have the calibration kit ready then press NEXT 3 Turn off the water supply Press Next There is a removable light shield attached ...

Страница 12: ...placement Kit To order any accessory or replacement parts please contact the HF scientific Customer Service Department HF scientific 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 Phone 239 337 2116 Toll Free 888 203 7248 Fax 239 454 0694 Email hf info wattswater com Website www hfscientific com 4 Disconnect the upper inlet tubing hose from the peristaltic pump and attach a single piece of...

Страница 13: ...gent tubing skip to step 16 12 Insert short piece of new tubing into outlet fitting on solenoid valve Insert other end into inlet fitting into optical injector 13 Secure solenoid valve to bracket with 2 screws removed in step 6 14 Insert end of new tubing into inlet of the solenoid valve 15 Reuse Velcro strap to secure the reagent tubing check valves to the top of the pump bracket 16 Guide tubing ...

Страница 14: ...the base 9 4 Replacing the Pressure Regulator Assembly Kit 28190S To replace the pressure regulator assembly on the CSM instrument 1 Turn off water to the CSM instrument 2 Loosen the screws on the bracket connecting the pressure regulator to the back panel 3 Lift bracket and attached components from back panel 4 Loosen the water inlet tube orange from the pressure regulator 5 Remove the panel nut ...

Страница 15: ...pump to the optical board and then remove air pump 11 Insert new air pump Be sure the arrow on the new air pump is facing the correct direction 12 Install 2 screws from underneath air pump bracket 13 Attach air pump to optical board using the included zip tie 14 Screw in black fitting into air pump outlet 15 Reconnect the black and pink air pump wiring connection located in the electrical panel 16...

Страница 16: ...IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE The remedy described in the first paragraph of this warranty shall constitute the sole and exclusive remedy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which ...

Отзывы: