10
5.11 Factory Reset
To reset all settings to factory defaults:
1. Press
Factory Reset
on the Configuration Menu screen.
2. Press
Reset
if you wish you to proceed with the factory reset. This
cannot be undone.
3. A status message will appear letting you know all configurable
parameters have been reset to factory defaults. A restart of the CSM
is then required.
5.12 About Device
The About Device menu displays information about the CSM.
To view any of this information:
1. Press
About Device
on the Configuration Menu screen. The
software version, software build date and hardware version will
be displayed.
2. Press
Cancel
to return to the previous menu.
6.0 System Start Up
Once all plumbing connections have been made, the CSM is turned on
and all configurations have been completed, it is ready for use.
6.1 Prepare Reagents
The reagent bag labels have a colored dot on them that corresponds
with the line it connects to on the CSM.
• Blue - Silver Buffer
• Yellow - Silver Indicator
• Red - Copper Indicator
• Green - Acid Digestion
To connect the reagents to the CSM:
1. Remove the cap from the reagent bag.
2. Gently squeeze out excess air from reagent bag, being careful not to
squeeze reagent liquid out of bag.
3. Connect the corresponding colored tube/cap to the reagent bag
with the same colored dot on the bag label.
4. Place the reagent bag onto the reagent bag bracket with the colored
tubing connector facing down in the CSM.
5. Repeat these steps for all 4 reagent bags.
6.2 Priming the CSM
Once the reagent bags have been properly connected, the CSM needs
to be primed before taking readings.
To prime the CSM:
1. From the
screen, press
to enter the Service Menu.
2. Press
PRIME
to enter the prime sub-menu.
3. Press the desired priming option. If this is the first time starting up
the instrument, start with
Prime Water
and repeat this step for all
priming options in this menu.
Reagent Bags Hanging on Brackets
P/N 110097
Assembled in the USA
Reagent Kit - Copper Indicator
16260 Airport Park Drive, Suite 140, Fort Myers, FL 33913
L# 74125 Rev E 11/2020
WARNING
Harmful if swallowed. Causes skin irritation. May cause an allergic skin reaction. Causes serious eye
irritation. Suspected of causing cancer. May cause damage to organs through prolonged or repeated
exposure.
Obtain special instructions before use. Do not handle until all safety precautions have been read
and understood. Do not breathe dust/fume/gas/mist/vapors/spray. Avoid breathing dust/fume/gas/mist/
vapors/spray. Wash hands, forearms and face thoroughly after handling. Do not eat, drink or smoke when
using this product. Contaminated work clothing should not be allowed out of the workplace. Wear protective
gloves/protective clothing/eye protection/face protection. IF SWALLOWED, call a POISON CENTER or doctor if
you feel unwell. IF ON SKIN, wash with plenty of water. IF IN EYES, rinse cautiously with water for several
minutes. Remove contact lenses, if present and easy to do. Continue rinsing. If exposed or concerned, get
medical advice/attention. Get medical advice/attention if you feel unwell. Rinse mouth. If skin irritation occurs,
get medical advice/attention. If skin irritation or rash occurs, get medical advice/attention. If eye irritation
persists, get medical advice/attention. Take off contaminated clothing and wash it before reuse. Store locked
up. Dispose of contents/container to hazardous or special waste collection point, in accordance with local,
regional, national and/or international regulation.
For use with CSM Instrument
Lot#
Exp XXX 20XX
Colored Dot
on Reagent
Bag Label
Reagent Bag Label