15.2.8 PROFIBUS GSD file
Qdos PROFIBUS pump can be integrated into PROFIBUS DP V0 network using a General Station Data (GSD) file.
File identifies pump and contains key data including:
l
Communication settings.
l
Commands it can receive.
l
Diagnostic information it can pass to PROFIBUS master on interrogation.
The GSD file—WAMA0E7D.GSD—can be either be:
l
Downloaded from Watson-Marlow website and installed.
l
Typed into PROFIBUS master directly using a GSD editor program.
NOTE52
Dataflow to/from pump may need to be byte-reversed, due to differences of handling data
between suppliers of master devices.
1
The GSD file, filename: WAMA0E7D.GSD
2
;
3
;************************************************************************************
4
;* ================================================================================ *
5
;* *
6
;* Watson-Marlow Bredel Pumps *
7
;* Bickland Water Road *
8
;* Falmouth *
9
;* Cornwall *
10
;* TR11 4RU *
11
;* Tel.: +44(1326)370370 *
12
;* FAX.: +44(1326)376009 *
13
;* *
14
;* ================================================================================ *
15
;* Filename: WAMA0E7D.GSD *
16
;* GSD file version 3 from 2013-09-24 *
17
;* -------------------------------------------------------------------------------- *
18
;* *
19
;************************************************************************************
20
#Profibus_DP
21
GSD_Revision = 3
22
Vendor_Name = “Watson Marlow”
23
Model_Name = “Qdos Profibus Pump”
24
Revision = “Version 3.00”
25
Ident_Number = 0x0E7D
26
Protocol_Ident = 0
27
Station_Type = 0
28
FMS_supp = 0
29
Hardware_Release = “V1.00”
30
Software_Release = “V1.00”
31
Redundancy = 0
32
Repeater_Ctrl_Sig = 0
33
24V_Pins = 0
34
9.6_supp = 1
35
19.2_supp = 1
36
45.45_supp = 1
15 INSTALLATION: PART 6 (SETTING UP THE PUMP (CONTROL SETTINGS))
153
Содержание qdos 20
Страница 40: ...6 Storage 6 1 Storage conditions 41 6 2 Shelf life 41 6 STORAGE 40...
Страница 169: ...Procedure 9 NEXT 10 Use keys to set Alert Level Procedure 11 SELECT to return to FLUID LEVEL SETTINGS 16 OPERATION 169...
Страница 192: ...NOTICE The pumphead locking lever is designed to be loosened or tightened by hand 17 MAINTENANCE 192...