Tip:
Use only the upper backing pump control connector. Leave the lower connector
marked Delayed backing pump control, disconnected.
3.
Connect the grounding cable to the grounding connection.
B.4 Connecting to the nitrogen gas supply
Required materials
• Chemical-resistant, powder-free gloves
• 6-mm PTFE tubing (included in the Xevo G2-S Tof Installation Kit)
• Nitrogen regulator (not supplied)
To connect the nitrogen gas supply:
1.
Connect one free end of the 6-mm PTFE tubing to the nitrogen inlet port on the rear of the
instrument.
Requirement:
Do not cut the 6-mm (¼-in) PTFE tubing to size; use the entire, supplied
5-m (16-ft) length.
Figure B–6: Nitrogen and collision cell gas inlet, and vacuum connectors:
Nitrogen exhaust
Nitrogen inlet port
Pilot valve port
Turbo vacuum