Setting Up the Calibration File
54
4
Chapter 4
Calibrating
Calibrating the mass scale entails setting up a calibration file and specifying calibration
parameters in MassLynx.
4.1 Setting Up the Calibration File
Before calibrating, you must remove the current calibration file and select a reference file:
1.
Select
Calibration > Calibrate Instrument
from the Tune window (
). The
Calibration window appears (
Figure 4-1 Calibration Window
2.
Select
NaICs2
from the reference file list to choose the reference file, if your ZQ
Mass Detector is a 2000 model. Select
NaICs4
if it is a 4000 model.
Reference File
List