background image

P

ROC

.# 715001121, R

EV

. A   R

EBUILDING

 

THE

 2487 A

UTOPURIFICATION

 F

LOW

 C

ELL

OF

 6

Rebuilding the 2487 Autopurification Flow Cell

This procedure outlines the steps necessary to rebuild the 2487 Autopurification Flow Cell. 

NOTE:

  To be certain you have the latest version of this procedure, please visit the Waters 

Support Center (http://www.waters.com).

Take the following precautions when disassembling and reassembling the flow cell:

• To prevent contamination, use powder-free finger cots or gloves when touching the

flow cell lenses or window.

• Take care to prevent scratching of the flow cell parts.
• Use a clean nonparticulating cloth or a similar clean surface in the work area where

you are removing, rebuilding, and replacing the flow cell.

NOTE:

  Always replace the flow cell gaskets when cleaning, rebuilding, or replacing other 

flow cell components.

Parts Required

Tools/Materials Required

• 1/4-inch flat-blade screwdriver
• Torque wrench set to 11 and 14 in-lb with 1/4-inch flat blade screwdriver bit
• Plastic tweezers
• Powder-free gloves or finger cots
• Lint-free swab
• Methanol

The cuvette lens mount screw, lens holder, entrance lens washer and lens mount will be reused 
when rebuilding the flow cell.

Table 1: Parts Required

Part Number

Description

Quantity

700002770

 2487 Autopurification Flow Cell Rebuild Kit

1

Table 2: Kit Components

Part Number

Description

Quantity

WAT081131

Gasket, Flow Cell

2

WAT081073

Lens, Cell Entrance

1

WAT081135

Split Ring, Cuvette

1

WAT081070

Lens, Cuvette

1

402000155

Window, Stepped, Flow Cell

1

Отзывы: