Operational error messages
5-5
Electronic A/D failure
Lamp optimization is
adjusted at the
minimum level.
Power-off and power-on
again.
Data acquisition via
A/D converters is
interrupt-driven. If
interrupt is too long,
problem with data
acquisition is indicated.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.
Filter initialization
failure: Erbium position
Unit sensors cannot
find erbium filter
position.
1. Ensure solvent is
flowing.
2. If the problem
persists, power-off
and power-on again.
3. If the problem
persists, call Waters
Technical Service.
Filter initialization
failure: Order filter
position
Unit sensors cannot
find the order filter
position.
1. Ensure solvent is
flowing.
2. If the problem
persists, power-off
and power-on again.
3. If the problem
persists, call Waters
Technical Service.
Filter initialization
failure: No filters found
Unit sensors observe
transition to dark
before homing the
optical filter.
1. Ensure solvent is
flowing.
2. If the problem
persists, power-off
and power-on again.
3. If the problem
persists, call Waters
Technical Service.
Error messages preventing operation (Continued)
Error message
Description
Corrective action
Содержание 2475
Страница 14: ...xiv Table of Contents...
Страница 158: ...4 16 Maintenance Procedures...
Страница 182: ...5 24 Error Messages Diagnostic Tests and Troubleshooting...
Страница 196: ...A 14 Safety Advisories...
Страница 200: ...B 4 Specifications...
Страница 214: ...C 14 Solvent Considerations...
Страница 228: ...Index 14...