3-50
Using the Detector
Specifying a method number
3. Specify a number from 1 through 10 and press Enter.
A brief message (“Storing * as method
n
”) appears, and then your
method number appears within the method icon. This method remains
active until you retrieve another method, or reset the detector to default
conditions (Method *).
Retrieving a method
To retrieve a method:
1. Select METHOD (Shift, A/B).
2. From the method list, select 3 Retrieve a method.
The last method number stored or retrieved appears in the method
number slot box.
3. Specify the number of the method you wish to retrieve and press Enter.
A brief message (“Retrieving method
n
”) appears, and then the method
number you specified appears within the method number icon (see the
table titled “2475 detector home and message screen icons” on page 3-5
).
Viewing events within a method
To view events within a method:
1. Retrieve the method (see
2. Press 1 to view the Timed events or 2 to view the Threshold events.
Tip:
If you change a timed or threshold event within a method, the
asterisk appears (Method *), indicating that the method (*) is no longer
the same as the stored method you retrieved in step 1. You can store the
method containing the altered event(s) in the same storage slot.