2-16
Installing the HPLC Pump
3. Use the 5/16-inch wrench to tighten the screw another quarter turn.
This seats the ferrule against the tubing.
4. Unscrew the assembled fitting, and verify that the length of tubing
extending beyond the ferrule is 3 mm.
5. Repeat
for the other end of the stainless steel tubing.
Performing the connections
To connect the outlet tubing to the injector or other device in your HPLC
configuration:
1. While pressing one end of the tubing assembly into the pump outlet
fitting, finger-tighten the compression screw, and then use the 5/16-inch
wrench to tighten the screw another 1/8-turn.
Tip:
Leave the instrument end of the outlet tubing disconnected until
you have primed the pump as described in
.
2. After you primed the pump, press the free end of the tubing assembly
into the injector or next device in your HPLC system, and finger-tighten
the compression screw.
3. Use the 5/16-inch wrench to tighten the screw another 1/8-turn.
Connecting fluid waste lines
Follow the instructions in this section to connect the fluid-waste lines. The
instructions describe how to do the following:
•
Connect the reference valve (or vent valve for 1525
μ
) and optional
manual injector waste lines
•
Connect the drip tray waste line
Required materials
•
Tubing (part number WAT076775 in startup kit)
•
Teflon tubing (part number WAT050693 in startup kit)
•
Convoluted ETFE tubing (part number WAT241095 in startup kit)
•
Plastic tubing cutter (not included) (part number WAT031795) or razor
blade
Содержание 1500-Series
Страница 14: ...xiv Table of Contents ...
Страница 32: ...1 18 Introduction ...
Страница 50: ...2 18 Installing the HPLC Pump ...
Страница 84: ...3 34 Installing Options and Accessories ...
Страница 108: ...4 24 Preparing for Operation ...
Страница 160: ...A 14 Safety Advisories ...
Страница 184: ...C 16 Solvent Considerations ...