2

Application Guidelines

(1)

Nominal product water ratings are based on the following conditions: Supply TDS of 250 ppm softened tap water, 50 psi (0.36 Mpa), 77

o

F (25

o

C), pH

8 and 15% recovery with outlet to atmosphere.

(2)

Rejection percentages are dependent on the supply conditions and the substance being measured.

Notes: 

The performance of a reverse osmosis membrane is highly dependent upon pressure, temperature and TDS. The actual volume of product
water and rejection percentage will vary with differences from the test conditions that membrane ratings are based upon.

These drinking water systems are not intended to be used for the treatment of water that is microbiologically unsafe or of unknown quality..

Functional Description

Feed  water  enters  the  5-micron  pre-filter  (combination  5  micron/carbon  filter  in  2  sump  models),  which
filters out suspended particles such as dirt or sediment. The filtered water then enters the pre-carbon filter
(3  sump  models  only),  which  contains  granular  activated  carbon,  which  removes  any  chlorine  from  the
water. The two sump models utilize a combination 5-micron/carbon filter to filter fine particles and remove
chlorine. 

The water then enters the reverse osmosis membrane. The membrane will allow only permeate (product
water) to pass through. The brine (waste water) goes to the drain.

Permeate then flows through a hydraulic shut-off valve to the storage tank. When the tank fills and the tank
pressure reaches 2/3 of the inlet feed water pressure, the shut-off valve closes, which turns the system
off.

When water is drawn from the faucet, permeate flows from the storage tank through the post carbon filter.
This filter contains granulated activated carbon, removing any taste and odor that has accumulated while
stored in the tank. When the tank empties, the shut-off valve opens, turning the unit on.

Model Number

E75TFC-3

E50TFC-3

E35TFC-2SF

Membrane Type

TFC

TFC

TFC

Max. Feed Water Salinity

2000 ppm

2000 ppm

2000 ppm

Feed Water Temperature

40-110°F

40-110°F

40-110°F

Feed Water Pressure

50-100 psi

50-100 psi

50-100 psi

Feed Water pH

2.0-11.0

2.0-11.0

2.0-11.0

Feed Water Supply

Chlorinated

Chlorinated

Chlorinated

Unchlorinated

Unchlorinated

Unchlorinated

Feed Water Hydrogen 
Sulphide

None

None

None

Feed Water Manganese

<0.05 ppm

<0.05 ppm

<0.05 ppm

Feed Water Iron

<0.1 ppm

<0.1 ppm

<0.1 ppm

Feed Water Hardness

<10 gpg

<10 gpg

<10 gpg

Specifications

Model Number

E75TFC-3SF

E50TFC-3SF

E35TFC-2SF

Membrane Production Rate 

(1)

75 USGPD

50 USGPD

35 USGPD

Rejection 

(1)

up to 99%

up to 99%

up to 99%

Storage Tank Capacity US Gal

4.0

4.0

4.0

Содержание E75TFC-3

Страница 1: ...Reverse Osmosis Drinking Water System Installation Operation and Service Manual 90012 Rev 06 05 June November 2005...

Страница 2: ......

Страница 3: ...h out the preservative solution before using the system This volume of water represents approximately two days of production The water flushed should be disposed of to the drain Feed Water Quality Thi...

Страница 4: ...ied items checklist 3 Installation 4 Startup and Operation 9 Maintenance 10 Sanitization Procedures 11 Maintenance Schedule 14 Parts List and Drawing for E75TFC 3SF E50TFC 3SF 15 Parts List and Drawin...

Страница 5: ...om the tap Ice Cubes use RO water to fill ice cube trays Ice cubes made from RO water are typically clearer and better tasting than ice made from plain tap water Automatic Ice Makers a water line from...

Страница 6: ...e reverse osmosis membrane The membrane will allow only permeate product water to pass through The brine waste water goes to the drain Permeate then flows through a hydraulic shut off valve to the sto...

Страница 7: ...ystem complete with tubing bundle 3 Reverse Osmosis membrane element in sealed protective bag 4 Filter Sump Wrench 5 Length of black tubing 6 Drain line saddle assembly 7 Feed Water saddle valve assem...

Страница 8: ...asily removed from under the sink Avoid placing the storage tank in out of the way cabinets or dead spaces There are two reasons for this 1 the difficulty of installation increases significantly and 2...

Страница 9: ...check for any signs of leaks f When ready to supply water to the system turn valve handle counter clockwise until fully open Feed Water Assembly Installation of Drain Saddle Assembly The system comes...

Страница 10: ...ap other than the one serving the garbage disposal Before making a connection as shown in Figure 5 please check with code authorities Figure 5 Drain Saddle Assembly Figure 6 Recommended Drain Saddle C...

Страница 11: ...sharp edges f Immediately clean up the sink area NOTE If an enamel or porcelain sink is encountered it is very important that the layer of porcelain be removed before following the steps outlined abo...

Страница 12: ...hand push the membrane all the way into the housing Important Do not use force and ensure that the membrane is not protruding from the housing 6 Replace the membrane housing end cap and hand tighten 7...

Страница 13: ...utlet side of the Pre Carbon Filter to the Auto Shut Off Valve Direct tubing into a pail or catch basin b Turn on the feed water by opening the needle valve on the feed water saddle valve and catch th...

Страница 14: ...of hydraulic pressure greater than the osmotic pressure in water containing dissolved solids The life of the membrane can be determined by measuring the percentage of rejection of total dissolved soli...

Страница 15: ...mponents DO NOT USE SCOURING PADS OR ANY OTHER TYPE OF INSTRUMENT THAT COULD POTENTIALLY SCRATCH THE SURFACES OF THE UNIT Measuring spoon or device Household bleach typically containing 5 25 of chlori...

Страница 16: ...SURFACE Unless the membrane element is being replaced at this time then discard the old membrane STEP 4 OPEN THE FILTER SUMPS WITH THE SUMP WRENCH SUPPLIED AND REMOVE THE CARTRIDGES STEP 5 CLEAN THE...

Страница 17: ...LVE STEP 10 ALLOW THE SYSTEM TO SIT WITH THE DILUTED SANITIZING SOLUTION FOR APPROXIMATELY 20 MINUTES STEP 11 OPEN THE FAUCET AND PURGE THE TANK AND SYSTEM OF THE SANITIZING SOLUTION STEP 12 REINSTALL...

Страница 18: ...edure Carbon Sediment 5 micron Sediment Carbon Cartridge 35 50 or 75 USGPD refer to Dual Cartridge Cartridge Rated Membrane instructions 2 sump models only 3 sump models 3 sump models x2 refer to mode...

Страница 19: ...404 14 Tee 3 8 x 1 4 npt 92403 16 Membrane RO 50 GPD TFC 92035 75 GPD TFC 92022 17 Housing cartridge white black 92026 18 Cartridge pre filter 5 micron 26091 19 Cartridge carbon 26081 20 Saddle drain...

Страница 20: ...t 92405 13 Elbow male 3 8 x 1 4 npt 92404 14 Swivel tee male 3 8 x 1 4 npt 92403 15 Connector union 3 8 1 4 92402 16 Membrane RO 35 GPD TFC 92008 17 Housing cartridge white black 92026 18b Cartridge c...

Страница 21: ...17 Reverse Osmosis Drinking Water System Flow Diagram all models...

Страница 22: ...in feed water TDS Contact dealer for revised product water TDS 3 Membrane life expired Replace membrane Bad tasting water 1 Taste from glass or plastic containers Contact dealer for explanation 2 Post...

Страница 23: ...19...

Страница 24: ...cteria attack installation or operation contrary to our printed instructions or damage caused by hot water freezing flood fire or Act of God is not covered by this warranty In all such cases regular p...

Отзывы: