COMMAND

USE

A = Set CAN address

Sets the CAN address

G = Select gas type

Select the gas type from a list

Z = Zero

Press when no gas on sensor to give zero

S = Span

     

 

H and L change reading

Use when calibration gas applied

D = Enter calibration date

Enter the calibration date

Y = Toggle auto zero

Auto zero is ON or OFF, small drift is cleared

H = Set high alarm

Sets the high alarm threshold

L = Set low alarm

Sets the low alarm threshold

O = Set over range alarm

Sets the over range alarm threshold

P = List command

List these commands on screen

X = Exit calibration mode

Exit this PC mode

$ = Initialise this sensor

Use on new PCB to set gas type to Flam

U = Alarm direction

Sets rising or falling alarms

R = Range

Allows a change in maximum value

N = Decimal points

Toggles between 1 and 2 decimal places

E = Edit user gas text

Choose gas description

B = Toggle deadband

Deadband of 2.5% can be on or off

F = Toggle fault Input

External fault input contact can be disabled

# = Normally energised

Low /high alarm relays and fault relay can be made normally energised

V = View gas log

From current log, display how many historical readings to display, up 

to 2880

% = Clear gas log

Set all 2880 log readings to 0.00

I = Log interval

Choose how many seconds between each log reading and whether the 

log will roll over or stop at 2880 (60 second interval and 2880 readings 

= 48 hours)

W = Set 4~20mA output

* Toggle Baud Rate 20kBits/40Kbits

USING MAGNETS (SET UP)

The  Combi  sensors  which  have  an  LCD  display  fitted  also  incorporate  3  reed  switches  which  can  be 

activated  using  external  magnets through the  glass window  of the flameproof XDIwin  enclosure. 

These 

magnets do not act instantly and have to be in close proximity to L, M and R on the front display for a few 

seconds to activate a software setup function.

The left magnet

 enters the Auto zero ON or OFF menu. This allows small drift changes in the sensor to 

be compensated for but is not operational when the sensor readings are greater than 5% of full scale. 

Therefore auto zero is inactive when a larger gas reading is present. When the remove magnets message 

appears, move the left magnet away and then the display shows if auto zero is ON or OFF. The left magnet 

puts auto zero on and the right magnet turns it off. With no magnets present, the display will return to 

normal after a few seconds time out.

16.  Pressing ‘SPACE BAR’ will exit the span mode

17.  Press ‘V’ to view log of sensor readings if required

18.  Pressing ‘X’ will exit the calibration mode.

NOTE:

 Oxygen cells only use the P+ and Y terminals J14. To adjust for “zero” it is normal practice to 

disconnect the green connector on J14 and adjust the Zero pot until LED D9 is green and 4mA across 

TP2/TP3. When the  cell  is  reconnected  in  air  at  20.9%  oxygen the  span  can  be  adjusted for 17.4mA 

reading across the same test pins using gain pot.

If an LCD option is fitted then calibration and other settings using magnets instead of a PC can be 

achieved – see over.

19.  On completion and when used as an addressable sensor, future calibrations may be carried out at the 

Combi alarm panel.