Initial Setting Mode

Initial Setting Mode

2. ACTIVATION 

2.1 SWITCH ON

Press and hold down the 

Key button

 (

) and the device will be switched on along with the three 

seconds countdown.

NOTE:

 

The device will be switched on only when you keep pressing the button for longer than 

three seconds.

                  

Once activated, the device will enter the warm up stage to stablize sensors. The warm up process 
is completed, the device is ready to detect gases.

CAUTION

A proper calibration is always required before using the device at the work 

site. The user shall check whether the device is properly sensing the levels of 

dangers of gases and make sure whether the detecting section of the device is 

not blocked with materials impairing the detection.

2.2 SWITCH OFF

Keep pressing the 

Key button

 (

) and the 3, 2 and 1 in the mentioned order will appear on the 

monitor and finally the device will be switched off.

NOTE:

 The device will not be switched off only unless you keep pressing the button for longer 

than three seconds.

Содержание QGM

Страница 1: ...USER MANUAL Portable multi gas detector QGM we watch gas where you can t...

Страница 2: ...ected concentrations 9 3 5 Check on Alarm Value 9 3 5 1 Initial Setting Concentration Levelsl 9 3 6 Dates and Time 10 3 7 Self Test 10 3 8 Check on bump test interval date and Latest bump test date 11...

Страница 3: ...o the environment such as temperature pressure and humidity Please make sure to cali brate the detector under the same or similar environment to the specification Extreme changes in temperature may ca...

Страница 4: ...sensor Dual CO H2S 4 Key 5 IR Port 6 Alarm LED 7 LCD display 8 Buzzer 1 PRODUCT OVERVIEW 4 3 7 1 2 DISPLAY SYMBOLS 8 6 6 6 6 High alarm Low alarm Alarm condition STEL Alarm TWA Alarm Fresh Air Calibr...

Страница 5: ...he device is ready to detect gases CAUTION A proper calibration is always required before using the device at the work site The user shall check whether the device is properly sensing the levels of da...

Страница 6: ...the levels exceed the threshold for either LOW alarm or HIGH alarm or TWA STEL the display icons of LOW HIGH TWA or STEL blinks regularly and the alarm LED and vibration activates When the device goes...

Страница 7: ...9 vol Acceptable hourly average exposure levels of the toxic gases for the last eight hours Time Weight Average Acceptable average exposure levels of the toxic gases for the last 15 minutes Short Term...

Страница 8: ...ust leave the area immediately and the sound alarm vibration LED alarm stops when the device goes to a safe area where the concentrations are normal TWA Alarm Sets Off The alarm sets off when the hour...

Страница 9: ...of the initialization 3 5 CHECK ON ALARM VALUE Press the Key button for three seconds under the ALARM VAL mode and the set value for the LOW alarm is displayed Press the Key button one time each to s...

Страница 10: ...monitoring of the supervisor The modification may be done through WatchGas IR LINK options 3 6 DATES AND TIME Press the button under the YY MM DD mode for 3 seconds and the day time mode will appear...

Страница 11: ...te via IR Link the set value for interval is displayed calibration interval date latest calibration date will be appeared in the mentioned order 4 EVENT LOG Up to 30 events may be saved and when the l...

Страница 12: ...n under the work environment where people may inhale gases 5 1 FRESH AIR CALIBRATION Press Key button for 3 seconds under the gas calibration value mode and the icon signifying fresh air calibration w...

Страница 13: ...the button it automatically enters the measure mode If the calibration fails the phrase FA Fail not OK appears on the LCD Press the button to enter the initial fresh air calibration mode and if you d...

Страница 14: ...econds In order to supply gases turn on the gas regulator Results should appear within roughly 20 seconds If the test is successful OK appears in all four corners of the display DOCKING STATION Standa...

Страница 15: ...ay LCD Backlight Indicator LED Alarm Visual LCD alarm display LCD Backlight Indicator LED Audible buzzer 90dB at 10cm Data Saving Event Log 30 EA Calibration Log 30 EA Bump Log 30EA Data log Two Month...

Страница 16: ...neglect or improper installation testing or calibrations Any unauthorized attempt to repair or modify the product or any other cause of damage beyond the range of the intended use including damage by...

Отзывы: