background image

Portable multi gas detector

QGM 

12

5. CALIBRATION

CAUTION

The  initial  calibration  is  executed  at  WatchGas  before  device  release.  The 

calibration values are saved in the device which means inaccurate calibration 

may impair the accuracy of the device performance. Normally, the calibration 

should be done once a year after the purchase and regularly every six months 

thereafter.

CAUTION

Because it is calibrated on the assumption that oxygen concentration is 

20.9%vol, the combustible gas is 0%LEL, and the toxic is 0ppm in the normal 

fresh atmosphere, fresh air calibration must be conducted in the absolutely 

clear air without any impact of other gases. Fresh air calibration in the airtight 

spaces therefore is not recommended. Make sure to avoid operation under the 

work environment where people may inhale gases.

5.1. FRESH AIR CALIBRATION

Press 

Key  button

 (

) for 3 seconds under the gas calibration value mode and the icon ( )      

signifying fresh air calibration will appear on the LCD monitor with the phrase "CAL ZERO". Press 

for another 3 seconds to do fresh air calibration and it takes 10 seconds to calibrate. Press the 

button during the calibration process to stop the calibration. If you press the button upon the 

completion, It will return to the fresh air calibration mode, and if you don't press the button, it 

automatically enters the measure mode.

If the calibration fails, FA (Fail), not OK appears on the LCD. Press the button to 

enter the initial fresh air calibration mode and it changes into the measure 

mode if you do not press the button for 3 seconds. If FA continues, please 

consult WatchGas or the store you purchased as it may require the replace-

ment of the sensor or repair of the device.

3 sec.

3 sec.

Содержание QGM

Страница 1: ...USER MANUAL Portable multi gas detector QGM we watch gas where you can t...

Страница 2: ...ected concentrations 9 3 5 Check on Alarm Value 9 3 5 1 Initial Setting Concentration Levelsl 9 3 6 Dates and Time 10 3 7 Self Test 10 3 8 Check on bump test interval date and Latest bump test date 11...

Страница 3: ...o the environment such as temperature pressure and humidity Please make sure to cali brate the detector under the same or similar environment to the specification Extreme changes in temperature may ca...

Страница 4: ...sensor Dual CO H2S 4 Key 5 IR Port 6 Alarm LED 7 LCD display 8 Buzzer 1 PRODUCT OVERVIEW 4 3 7 1 2 DISPLAY SYMBOLS 8 6 6 6 6 High alarm Low alarm Alarm condition STEL Alarm TWA Alarm Fresh Air Calibr...

Страница 5: ...he device is ready to detect gases CAUTION A proper calibration is always required before using the device at the work site The user shall check whether the device is properly sensing the levels of da...

Страница 6: ...the levels exceed the threshold for either LOW alarm or HIGH alarm or TWA STEL the display icons of LOW HIGH TWA or STEL blinks regularly and the alarm LED and vibration activates When the device goes...

Страница 7: ...9 vol Acceptable hourly average exposure levels of the toxic gases for the last eight hours Time Weight Average Acceptable average exposure levels of the toxic gases for the last 15 minutes Short Term...

Страница 8: ...ust leave the area immediately and the sound alarm vibration LED alarm stops when the device goes to a safe area where the concentrations are normal TWA Alarm Sets Off The alarm sets off when the hour...

Страница 9: ...of the initialization 3 5 CHECK ON ALARM VALUE Press the Key button for three seconds under the ALARM VAL mode and the set value for the LOW alarm is displayed Press the Key button one time each to s...

Страница 10: ...monitoring of the supervisor The modification may be done through WatchGas IR LINK options 3 6 DATES AND TIME Press the button under the YY MM DD mode for 3 seconds and the day time mode will appear...

Страница 11: ...te via IR Link the set value for interval is displayed calibration interval date latest calibration date will be appeared in the mentioned order 4 EVENT LOG Up to 30 events may be saved and when the l...

Страница 12: ...n under the work environment where people may inhale gases 5 1 FRESH AIR CALIBRATION Press Key button for 3 seconds under the gas calibration value mode and the icon signifying fresh air calibration w...

Страница 13: ...the button it automatically enters the measure mode If the calibration fails the phrase FA Fail not OK appears on the LCD Press the button to enter the initial fresh air calibration mode and if you d...

Страница 14: ...econds In order to supply gases turn on the gas regulator Results should appear within roughly 20 seconds If the test is successful OK appears in all four corners of the display DOCKING STATION Standa...

Страница 15: ...ay LCD Backlight Indicator LED Alarm Visual LCD alarm display LCD Backlight Indicator LED Audible buzzer 90dB at 10cm Data Saving Event Log 30 EA Calibration Log 30 EA Bump Log 30EA Data log Two Month...

Страница 16: ...neglect or improper installation testing or calibrations Any unauthorized attempt to repair or modify the product or any other cause of damage beyond the range of the intended use including damage by...

Отзывы: