Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
49
IM-0882-03
HoldReset
This is a function to cancel all holding state and display the latest display value during function ON.
The function becomes valid while the terminal is "0" level or shorted with the COM terminal.
DigitalZero
The digital zero function (DZ) is a function to set the display value to zero when the DZ instruction is made. The
display is offset during the DZ instruction, and the fluctuation width from the start of instruction is treated as
the display value.
"DZ" icon is displayed on the measurement screen during operation.
In addition, when restarting after turning off the power, it is also possible to set the offset by the DZ function at
last power off.
For details, refer to the contents of “D-ZeroRetention” in the system setting.
Caution
1
)
If the DZ instruction is issued during execution of various holding functions, the digital
zero function is executed immediately after the end of these holding.
2
)
If the DZ indication is issued during the display value is OVER or -OVER, the digital zero
is executed immediately after OVER or -OVER is lost.
MonitorChange
This is a function to switch the measurement display. The kind of displays are selected by [MENU]
⇒
[Display]
⇒
[DispSelect]
The screen is switched by shorting the terminal for 20 ms or longer with the COM terminal or setting it to "0"
level (falling edge detection), releasing the terminal once and shorting again will cause the screen to switch
again.
It is the same as the “DISP key” on the front panel.
TrendLog
This is a function to start acquiring trend data automatically apart from alarm log function. This function
operates even without alarm. Data storage starts with shorting the terminal for 20 ms or more or setting it to "0"
level (falling edge detection).
Please check the section about log for the contents of the log and how to view it.
* Log data cannot be saved even if trend log signal is input while trend data is being saved.
PatternChange
You can store and set up 8 patterns (8 kinds) of various settings such as input and output with this unit.
You can select a pattern by setting the terminals 1, 2, 3 to high or low.
Even if you only use one terminal, you can select pattern 1 or 2.
Display value
0
DZ: OFF
DZ: ON
External control
0
V
Short with COM
or
0 level