Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
22
IM-0882-03
8.
BASIC SETTINGS
CALIBRATION OF DC / PROCESS INPUT
Calibration of DC / process input is proceeded as follow.
Each setting items are set up at [Setting]
⇒
[Input]
⇒
[DC InputA/B] or [ProcessInputA/B].
When registering calibration values of multiple sensors, change the pattern and continue setting and
calibration.
Descriptions of each setting item are described on the following pages after this section.
Moreover, please check "input detailed setting" for the setting not described in the above flow.
*1 Each setting and calibration value are saved and reflected at the timing when the menu is closed and
returned to the measurement screen.
Note that the contents will not be saved if the power is turned off before closing the menu.
*If you changed the sensor power supply or
input range, please close the menu once for
reflecting the setting.
*For wiring, please confirm
"5. CONNECTING TERMINALS”.
Scaling
Pattern select
Sensor power supply
voltage select
Offset
calibration
Fullscale
calibration
Decimal point
position select
Unit select
Measurement start
Connect a sensor
Input range select