Instruction Manual for MODEL A9

11-0

□,

A9

12-0

 

5/8 

Example of setting scaling data: 
Example of setting 1 

Example of setting 2 
 

The digital limit function is a function for controlling display indication by 
concurrent  use  of digital limiter-type setting in the condition  data  and  a 
digital limit set value in the scaling data.     
If  CUT  is  selected  for  the  digital  limiter  type  in  the  condition  data,  the 
display value is held at the limit value set in the scaling data as shown in 
example of setting 2 above. Moreover, selection of OVER for the digital 
limiter  type  causes  O.L.  or 

–O.L.  to  appear  if  an  input  is  made  that 

results in exceeding the limit value set in the scaling data. 

 
 

4.8

 

Setting Linearize Data   

Linearize  data  is  a  group  of  parameters  relating  to  the  function  of 
correcting the  linearity  between input  and  display  values. The linearize 
function corrects the linear relationship between input and display values 
at any point to change the inclination of the linearity. Linearize data is set 
using  an input  value (display  value  before correction)  and  output  value 
(display value after correction) at any point.   
To use the linearize function, carry out this linearize data setup first and 
then  set  the  linearize  function  for  activation  in  the  condition  data.  The 
linearize function works only after that.   

 

 
(1) Press the Enter and Mode keys during measurement operation to 

  enter the condition data setting mode. 

 
(2) Press the Shift key twice to move to the linearize data setting 

  mode. 

 
(3) Press the Mode key to display linearize point number setting. 
 
(4)Set the number of linearize points using the Shift and Increment 
    keys. 

   

    (This setting example sets the number of linearize points to "10.") 
 

Shift key : Used to move to the setting digit. 
Increment key : Used to set a numeric value. 

 
* You must set the number of linearize points; otherwise you cannot   
move to the next item.   
 
 

 
(5) Press the Mode key to display a linearize point. 
 
(6) Press the Mode key to display input value setting with respect to         
the linearize point selected.   
"RE" LED also blinks during setting of an input value.   
*  The  input  value  is  a  display  value  for  an  input  before  executing 
linearization. 
 
(7) Press the Mode key to display output value setting with respect to 
the linearize point.   
"DZ" LED also blinks during setting of an output value.   
* The output value is a display value for the input made after 
execution of linearization. 

 

 

 
(8) Press the Mode key to display a next linearize point. 
*The input value and output value are similarly setting. 
  Please  set  to  be  suitable  for  the  following  setting  condition 
everything. 
(9) Press the Enter key to return to measurement operations.   
* After completion of setting, carry out linearize function setup in the 
condition data to activate this function for use. 
*  The  setting  conditions  must  be  N-01  <  N-02  ・・・  N-15  <  N-16, 
and  if  these  conditions  are  not  met,  ERR  appears.  If  this  happens, 
carry out the setting again. The number of linearize points is up to 16, 
but a value 17 to 19 is also displayed during setup. Note that if you 
set a value from 17 to 19, it is forced to set to 16. 

 

5. Other Functions   
5.1    Display Shift Function   

The  display  shift  function  is  a  function  for  arbitrarily  shifting  only  the 
indication without changing the inclination of an input signal. 

 
 
(1) Press the Enter and Mode keys during measurement operation to 
enter the display shift setting mode. 
 
 
 
(2) The panelmeter enters the shift numeric setting mode. 
 
 
(3)  Set  the  number  of  digits  by  which  you wish  to  shift  the  display 
indication, using the Shift and Increment keys. 
(This setting example sets a full-scale display value to "-345.") 

Shift  key:  Used  to  move  to  the  digit  where  you  carry  out  the 

setting 
    Increment key : Used to set a numeric value 
* The polarity is switched by incrementing the most significant digit. 
(4) Check the computation results using the Mode key.   

In this case, the decimal points of the 10

2

 digit and 10

3

 digit blink. 

 
 
(5) Press the Enter key to return to measurement operation.   

* To clear the display shift function, set "0." 

 
 
 

 

 

5.2    Monitoring Mode   

The A9000 series panelmeter can display the maximum value, minimum 
value, the difference between them (maximum value  - minimum value), 
or input  value in the  main  display. Pressing the Increment  key  with  the 
Enter key held down causes the panelmeter to enter the display status in 
each  mode.  To  switch  to  each  mode,  press  the  Shift  key  for 
approximately one second. This switches the display value in the order 
of  the  maximum  value,  minimum  value,  and  the  difference  (maximum 
value  -minimum  value),  and  the  input  value.  Moreover,  pressing  the 
Increment  key  for  approximately  one  second  allows  you  to  clear  the 
display  value.  Press  the  Enter  key  to  return  to  the  normal  indication. 
(The  next  time  you  enter  the  monitoring  mode,  the  mode  you  were  in 
when you exited on the previous occasion is activated.) 

Maximum value: The maximum value is displayed, blinking the decimal point of the 10

4

 digit.   

Minimum value:    The minimum value is displayed, blinking the 

  decimal point of the 10

0

 digit.   

Maximum value - minimum value:   

 

The difference between the maximum and minimum values is 

  displayed, blinking the decimal points of the 10

0

 and 10

4

 digits. If a 

display value exceeds the displayable range, the indication of the 10

3

 digit becomes “?”, lighting 

up its decimal point.   
Input value:    The input value is displayed, blinking the decimal points of the 10

0

 and 10

1

 digits. 

If a display value exceeds the displayable range, it displays O.L or –O.L. 
 

 

Displayed Value 

 

 

Input

 

120.00 

 

Input voltage:0 to 15V 

 

Display:  0 to 120.00

 

Select 13-range in condition 
  data setting

 

F S C : 

12000 

F I N :

 

15000

 

O F S :

 

0

 

O I N :

 

0

 

D L H I :

 

19999

 

D L L O :

 

-19999

 

D E P :

 

Lights  up  at  the 
10

²

  digit

 

 

[V]

 

20

 

15

 

10

 

5

 

 

E 

M 

▲ 

M 


 

(×twice) 


 

 

 


 

▲ 

M 

M 


 

E 

 

Displayed Value 

 

5.000 

 

2.500 

 

 

[mV]

 

Input

 

7.500 

 

Input voltage  :0 to 100mV 

 

Display            :0 to 5.000

 

Select  11-range  in  condition 
data setting,and choose CUT   
for the digital limiter type.

 

F S C : 

5000 

F I N :

 

10000

 

O F S :

 

0

 

O I N :

 

0

 

D L H I :

 

7500

 

D L L O :

 

-2500

 

D E P :

 

Light  up  at  the 
10

3

 digit.

 

 

-2.500 

 

200

 

100

 

DLLO 

 

DLHI 

 

 

M 

M 

M 

M 

M 

E 

Содержание A9000 Series

Страница 1: ...12 22 or 23 range Input terminal HI input terminal of 13 or 14 24 or 25 range Input terminal LO input terminal Selecting input range can not be done only by connecting input signal to a suitable termi...

Страница 2: ...vel 3 5 5V Input current 0 5mA 4 Parameter Settings 4 1 Components and Functions Before setting parameters remove the front panel by inserting a flat blade screwdriver into the ditch under the front p...

Страница 3: ...item 5 Press the Mode key to move to peak hold type setting 6 Set up the peak hold type using the Increment key This setting example sets peak hold type to PH PH Holds the maximum display value PV Ho...

Страница 4: ...Enter and Mode keys during measurement operation to enter the condition data setting mode 2 Press the Shift key to move to the scaling data setting mode 3 Press the Mode key to display the number of...

Страница 5: ...the setting again The number of linearize points is up to 16 but a value 17 to 19 is also displayed during setup Note that if you set a value from 17 to 19 it is forced to set to 16 5 Other Functions...

Страница 6: ...23 19 999mA 24 199 99mA 0 1 3A 25 1999 9mA Common specifications A D conversion conversion method Input circuit Single ended Setting sampling rate 25times sec max Overrange warning For an input signa...

Страница 7: ...ble ON disable OFF of the linearize function and data clear CLR TR T Tracking zero correction time 00 PL0 00 99 Sets the correction time of the tracking zero function Every TR T setting sampling perio...

Страница 8: ...period 1 25 times sec 40ms 20 1 25 times sec 800ms 2 12 5 times sec 80ms 40 0 625 times sec 1 6s 4 6 25 times sec 160ms 80 0 3125 times sec 3 2s 8 3 125 times sec 320ms 100 0 25 times sec 4s 10 2 5 ti...

Отзывы: